<strong>Sirona</strong> Dental Systems GmbH User interface in Master Mode<strong>Operator's</strong> <strong>Manual</strong>5.7.3.1.3 Occlusal offsetIf you find that all of your restorations have too little or too much material inthe occlusal direction, you can compensate for this with this parameter.The occlusal offset can be adjusted in increments of 25µm between -500µmand 500µm.ABIncreasing the occlusal offset (exaggerated representation)ABz offsetRestorationNOTICEVisible only on millingThis setting affects the restoration only on milling. The effect is not visible inthe milling preview.båÖäáëÜ5.7.3.1.4 Minimum thicknessWith these parameters you can set the desired minimum material thickness.This is also already visible during the design phase if the "Minimal thickness"checkbox is selected under "Window" / "Display options".The parameters "Minimal thickness (occlusal)" and "Minimal thickness (radial)"influence the minimum wall thickness geometry for veneers, inlays, onlaysand crowns only. Different values can be set for the occlusal and radialperspectives.59 56 458 D 3344D 3344.208.05.14.02 04.2011 61