Optical impressionOptical impressions with the scanner<strong>Sirona</strong> Dental Systems GmbH<strong>Operator's</strong> <strong>Manual</strong>Preparation model with bite registration1. Place preparation model A into the milling unit with bite registration Bmounted on top. The registration may cover the model over its entirelength.2. Have model A scanned using attached bite registration B as the antagonist.3. Then remove registration B from model A without removing the modelfrom the milling unit.4. Have the model scanned as the preparation.6.2.7 Scanning an occlusionWith this scanning technique, one scan is performed from the occlusal view(0°).The image field of the occlusion must agree with the image field of the preparationin the mesial and distal region (see also "General information on the opticalimpression").1. Insert the occlusion model in the milling unit.2. Have the occlusion scanned.3. Then remove the model from the milling unit.4. Insert the preparation model in the milling unit.5. Have the model scanned as the preparation.59 56 458 D 334488 D 3344.208.05.14.02 04.2011