Agilent Spectrum Analysis Basics - Agilent Technologies
Agilent Spectrum Analysis Basics - Agilent Technologies
Agilent Spectrum Analysis Basics - Agilent Technologies
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Figure 7-18. The image suppress function displays only real signals<br />
Note that both signal identification methods are used for identifying correct<br />
frequencies only. You should not attempt to make amplitude measurements<br />
while the signal identification function is turned on. Note that in both<br />
Figures 7-17 and 7-18, an on-screen message alerts the user to this fact.<br />
Once we have identified the real signal of interest, we turn off the signal<br />
ID function and zoom in on it by reducing the span. We can then measure<br />
the signal’s amplitude and frequency. See Figure 7-19.<br />
To make an accurate amplitude measurement, it is very important that you<br />
first enter the calibration data for your external mixer. This data is normally<br />
supplied by the mixer manufacturer, and is typically a table of mixer conversion<br />
loss, in dB, at a number of frequency points across the band. This data is<br />
entered into the ESA’s amplitude correction table. This table is accessed by<br />
pressing the [AMPLITUDE] key, then pressing the {More}, {Corrections},<br />
{Other} and {Edit} softkeys. After entering the conversion loss values, apply<br />
the corrections with the {Correction On} softkey. The spectrum analyzer<br />
reference level is now calibrated for signals at the input to the external mixer.<br />
If you have other loss or gain elements between the signal source and the<br />
mixer, such as antennas, cables, or preamplifiers, the frequency responses of<br />
these elements should be entered into the amplitude correction table as well.<br />
Figure 7-19. Measurement of positively identified signal<br />
101