05.08.2014 Views

Agilent Spectrum Analysis Basics - Agilent Technologies

Agilent Spectrum Analysis Basics - Agilent Technologies

Agilent Spectrum Analysis Basics - Agilent Technologies

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

For example, suppose that the LO fundamental has a peak-to-peak deviation<br />

of 10 Hz. The second harmonic then has a 20 Hz peak-to-peak deviation; the<br />

third harmonic, 30 Hz; and so on. Since the phase noise indicates the signal<br />

(noise in this case) producing the modulation, the level of the phase noise<br />

must be higher to produce greater deviation. When the degree of modulation<br />

is very small, as in the situation here, the amplitude of the modulation side<br />

bands is directly proportional to the deviation of the carrier (LO). If the<br />

deviation doubles, the level of the side bands must also double in voltage;<br />

that is, increase by 6 dB or 20 log(2). As a result, the ability of our analyzer<br />

to measure closely spaced signals that are unequal in amplitude decreases<br />

as higher harmonics of the LO are used for mixing. Figure 7-10 shows the<br />

difference in phase noise between fundamental mixing of a 5 GHz signal and<br />

fourth-harmonic mixing of a 20 GHz signal.<br />

Figure 7-10. Phase noise levels for fundamental and 4th harmonic mixing<br />

Improved dynamic range<br />

A preselector improves dynamic range if the signals in question have<br />

sufficient frequency separation. The discussion of dynamic range in Chapter 6<br />

assumed that both the large and small signals were always present at the<br />

mixer and that their amplitudes did not change during the course of the<br />

measurement. But as we have seen, if signals are far enough apart, a<br />

preselector allows one to reach the mixer while rejecting the others. For<br />

example, if we were to test a microwave oscillator for harmonics, a<br />

preselector would reject the fundamental when we tuned the analyzer<br />

to one of the harmonics.<br />

Let’s look at the dynamic range of a second-harmonic test of a 3 GHz<br />

oscillator. Using the example from Chapter 6, suppose that a –40 dBm signal<br />

at the mixer produces a second harmonic product of –75 dBc. We also know,<br />

from our discussion, that for every dB the level of the fundamental changes<br />

at the mixer, measurement range also changes by 1 dB. The second-harmonic<br />

distortion curve is shown in Figure 7-11. For this example, we shall assume<br />

plenty of power from the oscillator and set the input attenuator so that when<br />

we measure the oscillator fundamental, the level at the mixer is –10 dBm,<br />

below the 1 dB compression point.<br />

92

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!