This is the 2.5 dB factor that we accounted for in the previous preamplifier<br />

discussion, whenever the noise power out of the preamplifier was<br />

approximately equal to or greater than the analyzer’s own noise.<br />

Figure 5-7. The envelope of band-limited Gaussian noise has a Rayleigh distribution<br />

Another factor that affects noise measurements is the bandwidth in which<br />

the measurement is made. We have seen how changing resolution bandwidth<br />

affects the displayed level of the analyzer’s internally generated noise.<br />

Bandwidth affects external noise signals in the same way. To compare<br />

measurements made on different analyzers, we must know the bandwidths<br />

used in each case.<br />

Not only does the 3 dB (or 6 dB) bandwidth of the analyzer affect the<br />

measured noise level, the shape of the resolution filter also plays a role.<br />

To make comparisons possible, we define a standard noise-power bandwidth:<br />

the width of a rectangular filter that passes the same noise power as our<br />

analyzer’s filter. For the near-Gaussian filters in <strong>Agilent</strong> analyzers, the<br />

equivalent noise-power bandwidth is about 1.05 to 1.13 times the 3 dB<br />

bandwidth, depending on bandwidth selectivity. For example, a 10 kHz<br />

resolution bandwidth filter has a noise-power bandwidth in the range of<br />

10.5 to 11.3 kHz.<br />

If we use 10 log(BW 2 /BW 1 ) to adjust the displayed noise level to what we<br />

would have measured in a noise-power bandwidth of the same numeric value<br />

as our 3 dB bandwidth, we find that the adjustment varies from:<br />

10 log(10,000/10,500) = –0.21 dB<br />

to<br />

10 log(10,000/11,300) = –0.53 dB<br />

In other words, if we subtract something between 0.21 and 0.53 dB from the<br />

indicated noise level, we shall have the noise level in a noise-power bandwidth<br />

that is convenient for computations. For the following examples below, we<br />

will use 0.5 dB as a reasonable compromise for the bandwidth correction 9 .<br />

9. ESA Series analyzers calibrate each RBW during<br />

the IF alignment routine to determine the noise<br />

power bandwidth. The PSA Series analyzers<br />

specify noise power bandwidth accuracy to within<br />

1% (±0.044 dB).<br />