Chapter 4<br />

Amplitude and<br />

Frequency Accuracy<br />

Now that we can view our signal on the display screen, let’s look at amplitude<br />

accuracy, or perhaps better, amplitude uncertainty. Most spectrum analyzers<br />

are specified in terms of both absolute and relative accuracy. However,<br />

relative performance affects both, so let’s look at those factors affecting<br />

relative measurement uncertainty first.<br />

Before we discuss these uncertainties, let’s look again at the block diagram<br />

of an analog swept-tuned spectrum analyzer, shown in Figure 4-1, and see<br />

which components contribute to the uncertainties. Later in this chapter, we<br />

will see how a digital IF and various correction and calibration techniques<br />

can substantially reduce measurement uncertainty.<br />

RF input<br />

attenuator Mixer IF gain<br />

IF filter<br />

Log<br />

amp<br />

Envelope<br />

detector<br />

Input<br />

signal<br />

Pre-selector, or<br />

low-pass filter<br />

Local<br />

oscillator<br />

Video<br />

filter<br />

Reference<br />

oscillator<br />

Sweep<br />

generator<br />

Display<br />

Figure 4-1. <strong>Spectrum</strong> analyzer block diagram<br />

Components which contribute to uncertainty are:<br />

• Input connector (mismatch)<br />

• RF Input attenuator<br />

• Mixer and input filter (flatness)<br />

• IF gain/attenuation (reference level )<br />

• RBW filters<br />

• Display scale fidelity<br />

• Calibrator (not shown)<br />

1. For more information, see the <strong>Agilent</strong> PSA<br />

Performance <strong>Spectrum</strong> Analyzer Series Amplitude<br />

Accuracy Product Note, literature number<br />

5980-3080EN.<br />

An important factor in measurement uncertainty that is often overlooked<br />

is impedance mismatch. Analyzers do not have perfect input impedances,<br />

and signal sources do not have ideal output impedances. When a mismatch<br />

exists, the incident and reflected signal vectors may add constructively or<br />

destructively. Thus the signal received by the analyzer can be larger or<br />

smaller than the original signal. In most cases, uncertainty due to mismatch<br />

is relatively small. However, it should be noted that as spectrum analyzer<br />

amplitude accuracy has improved dramatically in recent years, mismatch<br />

uncertainty now constitutes a more significant part of the total measurement<br />

uncertainty. In any case, improving the match of either the source or<br />

analyzer reduces uncertainty 1 .<br />