Agilent Spectrum Analysis Basics - Agilent Technologies
Agilent Spectrum Analysis Basics - Agilent Technologies
Agilent Spectrum Analysis Basics - Agilent Technologies
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
In a factory setting, there is often an in-house frequency standard available<br />
that is traceable to a national standard. Most analyzers with internal<br />
reference oscillators allow you to use an external reference. The frequency<br />
reference error in the foregoing expression then becomes the error of the<br />
in-house standard.<br />
When making relative measurements, span accuracy comes into play.<br />
For <strong>Agilent</strong> analyzers, span accuracy generally means the uncertainty in<br />
the indicated separation of any two spectral components on the display.<br />
For example, suppose span accuracy is 0.5% of span and we have two signals<br />
separated by two divisions in a 1 MHz span (100 kHz per division). The<br />
uncertainty of the signal separation would be 5 kHz. The uncertainty would<br />
be the same if we used delta markers and the delta reading would be 200 kHz.<br />
So we would measure 200 kHz ±5 kHz.<br />
When making measurements in the field, we typically want to turn our<br />
analyzer on, complete our task, and move on as quickly as possible. It is<br />
helpful to know how the reference in our analyzer behaves under short warm<br />
up conditions. For example, the <strong>Agilent</strong> ESA-E Series of portable spectrum<br />
analyzers will meet published specifications after a five-minute warm up time.<br />
Most analyzers include markers that can be put on a signal to give us<br />
absolute frequency, as well as amplitude. However, the indicated frequency<br />
of the marker is a function of the frequency calibration of the display, the<br />
location of the marker on the display, and the number of display points<br />
selected. Also, to get the best frequency accuracy we must be careful to<br />
place the marker exactly at the peak of the response to a spectral component.<br />
If we place the marker at some other point on the response, we will get a<br />
different frequency reading. For the best accuracy, we may narrow the span<br />
and resolution bandwidth to minimize their effects and to make it easier to<br />
place the marker at the peak of the response.<br />
Many analyzers have marker modes that include internal counter schemes<br />
to eliminate the effects of span and resolution bandwidth on frequency<br />
accuracy. The counter does not count the input signal directly, but instead<br />
counts the IF signal and perhaps one or more of the LOs, and the processor<br />
computes the frequency of the input signal. A minimum signal-to-noise ratio<br />
is required to eliminate noise as a factor in the count. Counting the signal<br />
in the IF also eliminates the need to place the marker at the exact peak of<br />
the signal response on the display. If you are using this marker counter<br />
function, placement anywhere sufficiently out of the noise will do. Marker<br />
count accuracy might be stated as:<br />
±[(marker freq x freq ref error) + counter resolution]<br />
We must still deal with the frequency reference error as previously discussed.<br />
Counter resolution refers to the least significant digit in the counter readout,<br />
a factor here just as with any simple digital counter. Some analyzers allow<br />
the counter mode to be used with delta markers. In that case, the effects of<br />
counter resolution and the fixed frequency would be doubled.<br />
57