Let’s first test the two previous extreme cases.<br />

As NF PRE + G PRE – NF SA becomes less than –10 dB, we find that system noise<br />

figure asymptotically approaches NF SA – G PRE . As the value becomes greater<br />

than +15 dB, system noise figure asymptotically approaches NF PRE less 2.5<br />

dB. Next, let’s try two numerical examples. Above, we determined that the<br />

noise figure of our analyzer is 24 dB. What would the system noise figure be<br />

if we add an <strong>Agilent</strong> 8447D, a preamplifier with a noise figure of about 8 dB<br />

and a gain of 26 dB? First, NF PRE + G PRE – NF SA is +10 dB. From the graph<br />

of Figure 5-5 we find a system noise figure of about NF PRE – 1.8 dB, or about<br />

8 – 1.8 = 6.2 dB. The graph accounts for the 2.5 dB factor. On the other<br />

hand, if the gain of the preamplifier is just 10 dB, then NF PRE + G PRE – N FSA<br />

is –6 dB. This time the graph indicates a system noise figure of<br />

NF SA – G PRE + 0.6 dB, or 24 – 10 + 0.6 = 14.6 dB 7 . (We did not introduce<br />

the 2.5 dB factor previously when we determined the noise figure of the<br />

analyzer alone because we read the measured noise directly from the display.<br />

The displayed noise included the 2.5 dB factor.)<br />

Many modern spectrum analyzers have optional built-in preamplifiers<br />

available. Compared to external preamplifiers, built-in preamplifiers simplify<br />

measurement setups and eliminate the need for additional cabling. Measuring<br />

signal amplitude is much more convenient with a built-in preamplifier,<br />

because the preamplifier/spectrum analyzer combination is calibrated as a<br />

system, and amplitude values displayed on screen are already corrected for<br />

proper readout. With an external preamplifier, you must correct the spectrum<br />

analyzer reading with a reference level offset equal to the preamp gain. Most<br />

modern spectrum analyzers allow you to enter the gain value of the external<br />

preamplifier from the front panel. The analyzer then applies this gain offset<br />

to the displayed reference level value, so that you can directly view corrected<br />

measurements on the display.<br />

Noise as a signal<br />

So far, we have focused on the noise generated within the measurement<br />

system (analyzer or analyzer/preamplifier). We described how the measurement<br />

system’s displayed average noise level limits the overall sensitivity. However,<br />

random noise is sometimes the signal that we want to measure. Because of<br />

the nature of noise, the superheterodyne spectrum analyzer indicates a value<br />

that is lower than the actual value of the noise. Let’s see why this is so and<br />

how we can correct for it.<br />

By random noise, we mean a signal whose instantaneous amplitude has<br />

a Gaussian distribution versus time, as shown in Figure 5-6. For example,<br />

thermal or Johnson noise has this characteristic. Such a signal has no discrete<br />

spectral components, so we cannot select some particular component and<br />

measure it to get an indication of signal strength. In fact, we must define<br />

what we mean by signal strength. If we sample the signal at an arbitrary<br />

instant, we could theoretically get any amplitude value. We need some<br />

measure that expresses the noise level averaged over time. Power, which<br />

is of course proportionate to rms voltage, satisfies that requirement.<br />

7. For more details on noise figure, see <strong>Agilent</strong><br />

Application Note 57-1, Fundamentals of RF and<br />

Microwave Noise Figure Measurements, literature<br />

number 5952-8255E.<br />

65