We have already seen that both video filtering and video averaging reduce<br />

the peak-to-peak fluctuations of a signal and can give us a steady value.<br />

We must equate this value to either power or rms voltage. The rms value<br />

of a Gaussian distribution equals its standard deviation, σ.<br />

Figure 5-6. Random noise has a Gaussian amplitude distribution<br />

Let’s start with our analyzer in the linear display mode. The Gaussian noise<br />

at the input is band limited as it passes through the IF chain, and its envelope<br />

takes on a Rayleigh distribution (Figure 5-7). The noise that we see on<br />

our analyzer display, the output of the envelope detector, is the Rayleigh<br />

distributed envelope of the input noise signal. To get a steady value, the<br />

mean value, we use video filtering or averaging. The mean value of a Rayleigh<br />

distribution is 1.253 σ.<br />

But our analyzer is a peak-responding voltmeter calibrated to indicate the<br />

rms value of a sine wave. To convert from peak to rms, our analyzer scales<br />

its readout by 0.707 (–3 dB). The mean value of the Rayleigh-distributed<br />

noise is scaled by the same factor, giving us a reading that is 0.886 σ (l.05 dB<br />

below σ). To equate the mean value displayed by the analyzer to the rms<br />

voltage of the input noise signal, then, we must account for the error in the<br />

displayed value. Note, however, that the error is not an ambiguity; it is a<br />

constant error that we can correct for by adding 1.05 dB to the displayed<br />

value.<br />

In most spectrum analyzers, the display scale (log or linear in voltage)<br />

controls the scale on which the noise distribution is averaged with either<br />

the VBW filter or with trace averaging. Normally, we use our analyzer in the<br />

log display mode, and this mode adds to the error in our noise measurement.<br />

The gain of a log amplifier is a function of signal amplitude, so the higher<br />

noise values are not amplified as much as the lower values. As a result, the<br />

output of the envelope detector is a skewed Rayleigh distribution, and the<br />

mean value that we get from video filtering or averaging is another 1.45 dB<br />

lower. In the log mode, then, the mean or average noise is displayed 2.5 dB<br />

too low. Again, this error is not an ambiguity, and we can correct for it 8 .<br />

8. In the ESA and PSA Series, the averaging can be<br />

set to video, voltage, or power (rms), independent<br />

of display scale. When using power averaging, no<br />

correction is needed, since the average rms level is<br />

determined by the square of the magnitude of the<br />

signal, not by the log or envelope of the voltage.<br />

66