User Guide - Eurotherm Ltda

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

nanodac RECORDER/CONTROLLER: USER GUIDE4.17 STERILISER OPTIONThis block provides a means of recording complete sterilisation cycles, including for example, venting andpumping as well as the actual sterilising period. See Section 3.4.10 for display mode details.Data is stored in .uhh history files for viewing in Review software.SteriliserCycle statusRemainingEquilibrationSterilisingTotal CycleF 0 (A 0 )Running OutputPassed OutputStartStart 121°C121°C TimeStart 134°C134°C TimeTarget TimeCycle NumberAuto CounterFile by TagInput 1 TypePV1Target SPBand LowBand HighFailure DwellInput 2 TypeFailure DwellMeasured Temp.Target Temp.Z Temp.Low LimitWait Start00:00:0000:00:0000:00:0000:00:0000:00:00NoNoNoNo00:03:00No00:15:0000:03:000NoThermocouple013413413700:00:00Thermo r Detect00:00:0011513410134Figure 4.17 Steriliser block configuration menu4.17.1 Configuration parametersCycle Status Wait start: The cycle is waiting to be startedWaiting: Waiting for input 1 to reach its target setpoint.Equilibration: Currently in the equilibration periodSterilising: Currently in the sterilising phasePassed: The cycle has completed successfullyFailed: The cycle has failedTest cycle: A test cycle is in progressHA030554Issue 7 Nov 12Page 155