Calibration C-91. Connect the power supply.2. Connect a precision resistance box (called 6-dial in this manual) to theplatinum resistance thermometer input terminals.3. Turn the power ON.4. Move to the Calibration function group. This starts the 30-minute agingtimer. This timer provides an approximate timer for aging. After30 minutes, the No. 2 display changes to 0. You can advance to the nextstep in this procedure even if 0 is not displayed.Figure 3.205. Press the M key to display the count value for each input type. TheNo. 2 display at this time displays the currently entered count valueentered in Hexadecimal. Set the 6-dial as follows:• Input type 0: 390 Ω• Input type 1 or 3: 280 Ω• Input type 2 or 4: 140 ΩFigure 3.21Input type 0 Input types 1, 3 Input types 2, 46. Allow the count value on the No. 2 display to fully stabilize, then pressthe D key to temporarily register the calibration setup.7. Press the M key to set the Bulletin 900-TC32 to the state on the left. Setthe 6-dial as follows:• Input type 0: 10 Ω• Input type 1 or 3: 10 Ω• Input type 2 or 4: 100 ΩFigure 3.22Input type 0 Input types 1, 3 Input types 2, 48. Allow the count value on the No. 2 display to fully stabilize, then pressthe D key to temporarily register the calibration setup.9. Press the M key. The No. 2 display changes to the state on the left.Note that the data to be temporarily registered is not displayed when it isnot entirely prepared.Publication 900-UM003C-EN-E - January 2004

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!