Increased resolution for peptide mapping<br />
<strong>Column</strong>: ZORBAX 300SB-C18<br />
858750-902<br />
2.1 x 100 mm, 1.8 µm<br />
Mobile Phase: A: 0.1% TFA<br />
B: 0.01% TFA + 80% ACN<br />
Flow Rate: 0.5 mL/min<br />
Gradient: 2% B for 1 min, 2 to 45% B for 8.8 min,<br />
45 to 95% B for 0.2 min, 95% B for 2 min,<br />
95 to 2% B for 0.2 min<br />
Temperature: 50 ºC<br />
Detector: 1290 Infinity LC with diode array detector<br />
at 280 nm<br />
Sample: Enzymatic protein digest (MAb)<br />
Peptide Mapping <strong>Column</strong> <strong>Selection</strong><br />
Application Technique <strong>Agilent</strong> <strong>Column</strong>s Notes<br />
Large peptide<br />
fragments/hydrophobic<br />
peptide core<br />
Small hydrophobic<br />
peptides<br />
Biomolecule Separations<br />
400 bar HPLC Poroshell 300 SB-C18<br />
ZORBAX 300SB-C18, 3.5 µm<br />
<strong>Agilent</strong> 1200 LC<br />
600 bar UHPLC Poroshell 300 SB-C18 <strong>Agilent</strong> 1260 Infinity LC and 1260 Infinity Bio-inert Quaternary LC<br />
1200 bar UHPLC ZORBAX RRHD<br />
Poroshell 300 SB-C18, 1.8 µm<br />
<strong>Agilent</strong> 1290 Infinity LC<br />
400 bar HPLC Poroshell 120 EC-C18<br />
Poroshell 120 SB-C18<br />
<strong>Agilent</strong> 1200 LC<br />
600 bar UHPLC Poroshell 120 EC-C18<br />
Poroshell 120 SB-C18<br />
<strong>Agilent</strong> 1260 Infinity LC and 1260 Infinity Bio-inert Quaternary LC<br />
1200 bar UHPLC Poroshell 120 EC-C18<br />
Poroshell 120 SB-C18<br />
<strong>Agilent</strong> 1290 Infinity LC<br />
If you have an <strong>Agilent</strong> 1290 Infinity LC in your lab we recommend starting with a ZORBAX RRHD 300SB-C18 column to screen your peptide map.<br />
Norm<br />
350<br />
300<br />
250<br />
200<br />
150<br />
100<br />
50<br />
0<br />
-50<br />
1 2 3 4 5 6 7 8 9 min<br />
The longer 100 mm <strong>Agilent</strong> ZORBAX RRHD 300SB-C18 column provides maximum resolution for protein digests – in this sample the total run<br />
time, including washing and equilibration, is under fifteen minutes.<br />
Recommended column choices determined by system/column pressure maximum and peptide size/hydrophobicity.<br />
<strong>BioHPLC</strong> COLUMNS<br />
9