Increased resolution for peptide mapping<br />

<strong>Column</strong>: ZORBAX 300SB-C18<br />

858750-902<br />

2.1 x 100 mm, 1.8 µm<br />

Mobile Phase: A: 0.1% TFA<br />

B: 0.01% TFA + 80% ACN<br />

Flow Rate: 0.5 mL/min<br />

Gradient: 2% B for 1 min, 2 to 45% B for 8.8 min,<br />

45 to 95% B for 0.2 min, 95% B for 2 min,<br />

95 to 2% B for 0.2 min<br />

Temperature: 50 ºC<br />

Detector: 1290 Infinity LC with diode array detector<br />

at 280 nm<br />

Sample: Enzymatic protein digest (MAb)<br />

Peptide Mapping <strong>Column</strong> <strong>Selection</strong><br />

Application Technique <strong>Agilent</strong> <strong>Column</strong>s Notes<br />

Large peptide<br />

fragments/hydrophobic<br />

peptide core<br />

Small hydrophobic<br />

peptides<br />

Biomolecule Separations<br />

400 bar HPLC Poroshell 300 SB-C18<br />

ZORBAX 300SB-C18, 3.5 µm<br />

<strong>Agilent</strong> 1200 LC<br />

600 bar UHPLC Poroshell 300 SB-C18 <strong>Agilent</strong> 1260 Infinity LC and 1260 Infinity Bio-inert Quaternary LC<br />

1200 bar UHPLC ZORBAX RRHD<br />

Poroshell 300 SB-C18, 1.8 µm<br />

<strong>Agilent</strong> 1290 Infinity LC<br />

400 bar HPLC Poroshell 120 EC-C18<br />

Poroshell 120 SB-C18<br />

<strong>Agilent</strong> 1200 LC<br />

600 bar UHPLC Poroshell 120 EC-C18<br />

Poroshell 120 SB-C18<br />

<strong>Agilent</strong> 1260 Infinity LC and 1260 Infinity Bio-inert Quaternary LC<br />

1200 bar UHPLC Poroshell 120 EC-C18<br />

Poroshell 120 SB-C18<br />

<strong>Agilent</strong> 1290 Infinity LC<br />

If you have an <strong>Agilent</strong> 1290 Infinity LC in your lab we recommend starting with a ZORBAX RRHD 300SB-C18 column to screen your peptide map.<br />

Norm<br />

350<br />

300<br />

250<br />

200<br />

150<br />

100<br />

50<br />

0<br />

-50<br />

1 2 3 4 5 6 7 8 9 min<br />

The longer 100 mm <strong>Agilent</strong> ZORBAX RRHD 300SB-C18 column provides maximum resolution for protein digests – in this sample the total run<br />

time, including washing and equilibration, is under fifteen minutes.<br />

Recommended column choices determined by system/column pressure maximum and peptide size/hydrophobicity.<br />

<strong>BioHPLC</strong> COLUMNS<br />