<strong>Column</strong> Length<br />

size exclusion Chromatography (seC)<br />

Where the separation time is a critical parameter shorter columns packed with the higher efficiency 3 µm media are used. With the shorter<br />

columns higher flow rates are used to reduce the analysis time but without compromising the quality of the data – quantitation of monoclonal<br />

antibody monomer and dimer.<br />

<strong>Agilent</strong> Bio SEC-3 column length comparison, 150 mm<br />

<strong>Column</strong>: Bio SEC-3, 300Å<br />

5190-2512<br />

7.8 x 150 mm, 3 µm<br />

Mobile Phase: 150 mM sodium phosphate<br />

Flow Rate: 1.0 mL/min (56 bar), 1.5 mL/min (75 bar)<br />

Detector: 220 nm<br />

Sample: MAb (2 mg/mL)<br />

150 mm 300 mm<br />

1.5 mL/min 1.0 mL/min<br />

Monomer<br />

Dimer<br />

Monomer<br />

Dimer<br />

4 min<br />

<strong>Agilent</strong> Bio SEC-3 column length comparison, 300 mm<br />

<strong>Column</strong>: Bio SEC-3, 300Å<br />

5190-2511<br />

7.8 x 300 mm, 3 µm<br />

Mobile Phase: 150 mM sodium phosphate + 100 mM Na sulfate (with salt)<br />

150 mM sodium phosphate (without salt)<br />

Flow Rate: 1.0 mL/min<br />

Detector: 220 nm<br />

Sample: MAb (2 mg/mL)<br />

With salt<br />

Without salt<br />

Monomer<br />

Dimer<br />

8 min<br />