BioHPLC Column Selection Guide Cover - Agilent Technologies
BioHPLC Column Selection Guide Cover - Agilent Technologies
BioHPLC Column Selection Guide Cover - Agilent Technologies
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
MicroBore (1.0 mm id) <strong>Column</strong>s<br />
Separation of a tryptic digest<br />
on ZORBAX MicroBore 300SB-C18<br />
<strong>Column</strong>: ZORBAX 300SB-C18<br />
863630-902<br />
1.0 x 150 mm, 3.5 µm<br />
Mobile Phase: Gradient: 2-60% B in 60 min<br />
A: 0.1% TFA<br />
B: 0.075% TFA/80% ACN<br />
Flow Rate: 50 µL/min<br />
Temperature: 50 °C<br />
Detector: 215 nm<br />
Sample: 2 µL<br />
Tryptic Digest of rhGH<br />
This example of a tryptic digest separated on a<br />
MicroBore column demonstrates the high sensitivity<br />
and resolution possible with 1.0 mm id columns.<br />
116 www.agilent.com/chem/biohplc<br />
MICROBORE (1.0 MM id) COLUMNS<br />
• High sensitivity for small sample sizes<br />
• Compatible with LC/MS interfaces<br />
• Wide variety of bonded phases<br />
• Silica and polymeric particles<br />
<strong>Agilent</strong> MicroBore (1.0 mm id) columns are a good choice when sample sizes are limited. They can<br />
improve detection limits 5 times over 2.1 mm id columns when the same sample mass is used. This<br />
increase in sensitivity can be critical. MicroBore columns use low flow rates (typically ~ 50 µL/min).<br />
Therefore, these columns are ideal for use with detectors requiring low flow rates such as some mass<br />
spectrometers and with capillary LC systems.<br />
Optimum performance is achieved when MicroBore columns are used with UHPLC/HPLC<br />
Microbore systems. A wide variety of bonded phases is available for use up to 400 bar including<br />
StableBond, 300SB-C18, 300SB-C8, and Poroshell columns. Polymeric reversed-phase, PLRP-S,<br />
and ion-exchange PL-SAX and PL-SCX are also available for applications requiring exceptionally<br />
stable wide pore particles. Guard columns are also now available with an adjustable tube stop<br />
depth to provide a perfect zero dead volume connection every time.<br />
0 5<br />
10 15 20 25 30 35 40 45 50 55<br />
Time (min)<br />
LCMB001