Improved reproducibility of monoclonal antibodies<br />
<strong>Column</strong>: ZORBAX RRHD 300SB-C8<br />
857750-906<br />
2.1 x 50 mm, 1.8 µm<br />
Mobile Phase: A: H20:IPA (98.2), 0.1% TFA<br />
B: IPA:ACN:H20 (70:20:10), 0.1% TFA<br />
Flow Rate: 1.0 mL/min<br />
Temperature: 80 ºC<br />
Detector: 1290 Infinity LC with diode array detector at 225 nm<br />
Sample: MAb<br />
Reversed-Phase HPLC<br />
0 0.5 1 1.5 2 2.5 3 min<br />
Gradient timescale<br />
Time (min) % Solvent B<br />
500<br />
pressure<br />
gradient wash and<br />
re-equilibration region<br />
0.00 25<br />
300<br />
3.00 35<br />
4.00 90<br />
100<br />
UV<br />
5.00 25 0 5 min<br />
Excellent column reproducibility and protein recovery using <strong>Agilent</strong> ZORBAX 300SB-C8.<br />
Increased resolution for peptide mapping<br />
<strong>Column</strong>: ZORBAX 300SB-C18<br />
858750-902<br />
2.1 x 100 mm, 1.8 µm<br />
Mobile Phase: A: 0.1% TFA<br />
B: 0.01% TFA + 80% ACN<br />
Flow Rate: 0.5 mL/min<br />
Gradient:<br />
Temperature: 50 ºC<br />
2% B for 1 min, 2 to 45% B<br />
for 8.8 min, 45 to 95% B<br />
for 0.2 min, 95% B for 2 min,<br />
95 to 2% B for 0.2 min<br />
Detector: 1290 Infinity LC with diode array<br />
detector at 280 nm<br />
Sample: Enzymatic protein digest (MAb)<br />
Norm<br />
350<br />
300<br />
250<br />
200<br />
150<br />
100<br />
50<br />
0<br />
-50<br />
mA<br />
400<br />
300<br />
200<br />
100<br />
mA<br />
700<br />
1st injection<br />
150th injection<br />
pre- and post-150<br />
injection column<br />
blank run overlays<br />
1 2 3 4 5 6 7 8 9 min<br />
The longer 100 mm <strong>Agilent</strong> ZORBAX RRHD 300SB-C18 column provides maximum resolution for protein digests – in this sample the total run<br />
time, including washing and equilibration, is under five minutes.<br />
bar<br />
750<br />
500<br />
250<br />
UHPLC/HPLC<br />
2