BioHPLC Column Selection Guide Cover - Agilent Technologies
BioHPLC Column Selection Guide Cover - Agilent Technologies
BioHPLC Column Selection Guide Cover - Agilent Technologies
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
LC/MS analysis of angiotensin<br />
on Extend-C18<br />
<strong>Column</strong>: ZORBAX Extend-C18<br />
773700-902<br />
2.1 x 150 mm, 5 µm<br />
Mobile Phase: Acidic Conditions:<br />
A: 0.1% TFA in water<br />
B: 0.085% TFA in 80% acetonitrile (ACN)<br />
Basic Conditions:<br />
A: 10 mM NH4OH in water<br />
B: 10 mM NH4OH in 80% ACN<br />
A<br />
Angiotensin I<br />
Max: 10889<br />
Low pH<br />
B<br />
Angiotensin I<br />
Max: 367225<br />
High pH<br />
TIC (200-1500 m/z)<br />
TIC (200-1500 m/z)<br />
5.0E6<br />
4.0E6<br />
3.0E6<br />
2.0E6<br />
1.0E6<br />
0<br />
5.0E7<br />
4.0E7<br />
3.0E7<br />
2.0E7<br />
1.0E7<br />
0<br />
0<br />
0<br />
Flow Rate: 0.2 mL/min<br />
Gradient: 15-50% B in 15 min<br />
Temperature: 35 °C<br />
2.5<br />
2.5<br />
AII + AIII<br />
5.0 7.5 10 12.5<br />
Time (min)<br />
AII<br />
5.499<br />
AI<br />
AIII<br />
8.477<br />
9.621<br />
12.050<br />
13.261<br />
5.0 7.5 10 12.5<br />
Time (min)<br />
AI<br />
100<br />
80<br />
60<br />
40<br />
20<br />
0<br />
0<br />
+3<br />
227.1<br />
250.9<br />
340.3<br />
398.0<br />
433.0<br />
458.0<br />
433.0<br />
Reversed-Phase HPLC<br />
Both small and large peptides demonstrate selectivity changes at high and low pH. At high pH, due to a change in charge, all three Angiotensins<br />
can be resolved. In addition, the spectral clarity of Angiotensin I is dramatically improved at high pH with the ammonium hydroxide mobile phase.<br />
The Extend-C18 column can be used for the analysis of small peptides at high pH as well.<br />
Reference: B.E. Boyes. Separation and Analysis of Peptides at High pH Using RP-HPLC/ESI-MS, 4th WCBP, San Francisco, CA, Jan. 2000.<br />
MS Conditions: Pos. Ion ESI- Vf 70 V, Vcap 4.5 kV,<br />
N2- 35 psi, 12 L/min., 325 °C<br />
Sample: 2.5 µL sample (50 pmol each)<br />
Angiotensin I, II, III<br />
100<br />
80<br />
60<br />
40<br />
20<br />
0<br />
0<br />
500<br />
+3<br />
500<br />
649.5<br />
+2<br />
659.7<br />
720.3<br />
767.3<br />
844.8<br />
649.0<br />
659.8<br />
+2<br />
1025.2<br />
1000<br />
1000<br />
1179.1<br />
1277.9<br />
1326.3<br />
1373.3<br />
1296.6<br />
1468.8<br />
m/z<br />
+1<br />
m/z<br />
LC30003<br />
UHPLC/HPLC<br />
2