Poroshell 120<br />
• 120Å pore size for shorter chain peptide mapping<br />
• UHPLC performance on 600 bar systems<br />
• Up to 90% of the efficiency of sub-2 µm<br />
• 2X the efficiency of 3.5 µm<br />
• Up to 50% less pressure than sub-2 µm columns<br />
Reversed-Phase HPLC<br />
<strong>Agilent</strong> Poroshell 120 columns are a 2.7 µm particle with a 1.7 µm solid core and 0.5 µm porous<br />
outer layer. This small particle size provides high efficiency, similar to sub-2 µm columns, but with<br />
40-50% less pressure. These high efficiency, high resolution columns can be used on any type of LC.<br />
The porous outer layer and solid core limit diffusion distance and improve separation speed while<br />
the narrow particle size distribution improves efficiency and resolution. The columns can support<br />
high pressure and multiple columns can be used for the highest resolution and efficiency possible.<br />
The smaller 120Å pore size is ideal for fast high resolution analysis of small hydrophilic peptides<br />
in protein digests.<br />
<strong>Column</strong> Specifications<br />
Bonded Phase Pore Size Temp. Limits pH Range Endcapped<br />
Carbon<br />
Load<br />
SB-C18 120Å 90 °C 1.0-8.0 No 8%<br />
EC-C18 120Å 60 °C 2.0-8.0 Double 10%<br />
Specifications represent typical values only.<br />