Analysis of representative whey proteins<br />
<strong>Column</strong>: PL-SAX 1000Å<br />
PL1551-1802<br />
4.6 x 50 mm, 8 µm<br />
Mobile Phase: A: 0.02 M Tris HCI, pH 7<br />
B: A + 0.5 M CH3COONa, pH 7<br />
Flow Rate: 1.0 mL/min<br />
Gradient: Linear 0-50% B in 10 min<br />
Detector: UV, 280 nm<br />
High resolution separation of<br />
a Poly-T-Oligonucleotide size standard<br />
spiked with 10mer, 15mer, 30mer<br />
and 50mer (main peaks)<br />
<strong>Column</strong>: PL-SAX 1000Å<br />
PL1551-1802<br />
4.6 x 50 mm, 8 µm<br />
Mobile Phase: A: 7:93 v/v ACN: 0.1 M TEAA, pH 8.5<br />
B: 7:93 v/v ACN: 0.1 M TEAA,<br />
1 M ammonium chloride, pH 8.5<br />
Gradient: 0-40% B in 10 min, followed by 40-70% B<br />
in 14 min and 70-100% B in 25 min<br />
Flow Rate: 1.5 mL/min<br />
Temperature: 60 °C<br />
Detector: UV, 220 nm<br />
2000<br />
0<br />
0<br />
1<br />
0<br />
2<br />
3<br />
4<br />
ion-exchange Chromatography<br />
min 18<br />
10<br />
15<br />
1. Carbonic anhydrase<br />
2. α-lactalbumin<br />
3. β-lactoglobulin B<br />
4. β-lactoglobulin A<br />
30 50<br />
min 50<br />
VLC0045<br />
UHPLC/HPLC<br />
5