Advanced Coupler Tests 93<br />

It then continues through a complete cycle and terminates when it reaches the<br />

baseline again. This signal, although consisting of only a single pulse, contains<br />

very few frequencies above 1 kHz and is one millisecond wide at its base.<br />

The data collected from this measurement is displayed in a graphical format 20<br />

milliseconds wide. The system delay and the digital processing delay are noted<br />

for reference on the phase graph in the form of vertical lines.<br />

4.1.3 Procedure<br />

1. Press [F3] in the Opening screen to enter the Advanced Coupler Navigation<br />

screen.<br />

2. Press [F4] in the Advanced Coupler Navigation screen to enter the<br />

Enhanced DSP screen.<br />

3. Set up the hearing aid for testing in the sound chamber. See Section 3.3.<br />

Make sure the sound chamber is leveled.<br />

4. Use [F1] to select the desired ear.<br />

Figure 4.1.3—Enhanced DSP measurement results<br />

5. Press [START] to perform the delay and phase tests for that hearing aid. If<br />

the left ear was designated in Step 4, the results will be shown in the upper<br />

two graphs. Otherwise, they will be shown in the lower two graphs.