70 FONIX <strong>8000</strong> Hearing Aid Test System<br />

Figure 3.4.6—Measuring a three frequency average<br />

3.4.7 Measuring Harmonic Distortion<br />

Harmonic distortion is when the hearing aid returns energy at frequencies not<br />

present in the source signal. The FONIX <strong>8000</strong> Test System allows you test for<br />

this phenomenon very easily.<br />

1. Press [MENU] from the Coupler screen.<br />

2. Select Distortion under Measurement Settings using the [, ] buttons.<br />

3. Select the distortion type desired by using the [, ] buttons.<br />

The selections are:<br />

2ND: Distortion present at the second harmonics, or twice the presented<br />

frequency.<br />

3RD: Distortion present at the third harmonics, or three times the presented<br />

frequency.<br />

TOTAL: Distortion present at the second and third harmonics.<br />

4. Press [EXIT].<br />

5. Use [F5] to select a signal type of TONE NORMAL. (Press [F5]. Use [, ]<br />

to select TONE NORMAL. Use [] to complete the selection and close the<br />

pop-up menu.)<br />

6. Press [START] to run normal frequency sweep.