02.08.2013 Views

FONIX® 8000 - Frye Electronics

FONIX® 8000 - Frye Electronics

FONIX® 8000 - Frye Electronics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

38 FONIX <strong>8000</strong> Hearing Aid Test System<br />

2.3.2.1 Noise reduction<br />

Composite noise reduction is a little different than pure-tone noise reduction<br />

(discussed in Section 2.3.1.1), even though both are used for noisy testing environments.<br />

When a composite signal is running, the analyzer takes several different measurements<br />

a second and displays them on the screen. Composite noise reduction<br />

performs a “running average” of these composite measurements. This<br />

means it averages together several of the previous measurements with the current<br />

measurement to produce the next curve. If you select “2X” noise reduction,<br />

it will average the last two measurements together. A selection of “4X”<br />

averages the last four measurements together. When performing composite measurements,<br />

it will take several seconds for the actual noise reduction used to<br />

reach the set value. This is due to the nature of the running average.<br />

Larger noise reduction numbers lead to smoother curves but increase the<br />

amount of time it takes the analyzer to update its composite measurements.<br />

2.3.2.2 Intermodulation distortion<br />

The composite signals are helpful for identifying intermodulation distortion<br />

(IM). IM distortion occurs when amplitudes at more than one frequency in a<br />

signal combine to create an amplitude at a frequency not present in the original<br />

signal. When viewing a graph run with a composite signal, look for points along<br />

the graph where the line “breaks up.” Such an appearance indicates the presence<br />

of IM distortion. See Figure 2.3.2.2 for an example of IM distortion.<br />

This type of distortion is only apparent when a composite signal source is used<br />

because pure-tone sweeps do not present more than one frequency at a time.<br />

Figure 2.3.2.2—Example of IM distortion

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!