168 FONIX <strong>8000</strong> Hearing Aid Analyzer<br />

Most of the operations of the Real-ear SPL screen are identical to the operations of<br />

the Insertion Gain screen. Therefore, most of the instructions found in this section<br />

apply to measurements performed in both screens. Any differences between operations<br />

in the two screens are noted.<br />

To enter the Real-ear SPL screen, press [F2] from the Opening screen to enter the<br />

Real-ear Navigation screen. Then press [F4].<br />

Figure 8.4.2—Rear-ear SPL screen. The T curve represnts the patient’s thresholds, the L, M, and H<br />

curves represent the NAL-NL1 targets. Curve 1 shows the unaided response. Curves 2, 3, and 4 show<br />

the aided response at 50, 65, and 80 dB SPL. The patient’s UCLs are just off the top of the graph.<br />

8.4.3 Measuring the Unaided Response (REUG/REUR)<br />

The REUG/REUR is a measurement of the unaided response of the patient’s ear<br />

canal. It can be measured in either the Real-ear Insertion Gain screen or the Realear<br />

SPL screen, and any measurement performed in one of these screens will be<br />

automatically displayed in the other screen.<br />

to measure the unaided response:<br />

1. Position the patient in front of the sound field speaker and place the earhook and<br />

reference microphone on the ear. See Sections 8.2.1 and 8.2.2.<br />

2. Insert the probe tube into the patient’s ear canal. See Section 8.2.3.<br />

3. Input the patient’s audiogram to generate a target, if desired. See Section 8.3.<br />

4. Press [EXIT] in the Audiogram Entry screen to return to the Real-ear Navigation<br />

screen.