84 FONIX <strong>8000</strong> Hearing Aid Test System<br />

8. Press [START] to perform the measurement. You should see CIC appear in<br />

the Curve Characteristics box in the “Cor” column of the selected curve. If<br />

you are using the FAST, COMPOSITE, or DIG SPEECH signal, press [STOP]<br />

to end the measurement once it has stabilized.<br />

Figure 3.6 shows the difference between a CIC hearing aid tested with an HA-1<br />

coupler, and the same aid tested with a CIC coupler and correction factors.<br />

Figure 3.6— CURVE 1 uses CIC coupler + correction factors<br />

CURVE 2 uses 2-cc HA-1 coupler<br />

3.7 Automatic Testing—Loading & Saving Groups<br />

You can use the FONIX <strong>8000</strong> Test System to program simple automated or semiautomated<br />

test sequences that will help you test hearing aids more efficiently.<br />

Each test sequence can contain up to 10 different response curves with your<br />

choice of source types, source levels, noise reduction, and other adjustments.<br />

In automatic mode, the analyzer will quickly run through all designated tests<br />

without pausing. In semi-automatic mode, the analyzer will automatically<br />

advance to the next curve and stop. This curve will not be measured until you<br />

start the test, but you are saved the step of using F2 to select the next curve<br />

number.