140 FONIX <strong>8000</strong> Hearing Aid Test System<br />

Figure 6.5.1<br />

6.5.2 Testing in Simultaneous Mode<br />

By default, when you first enter the Directional Difference test screen, you are<br />

in Simultaneous Mode. In Simultaneous Mode, all five difference curves are<br />

taken in one test sequence. A single zero degree “forward” measurement is<br />

used when calculating all five difference curves taken with potentially different<br />

reverse angles.<br />

To test in Simultaneous Mode:<br />

1. Use the [F6] key, if necessary, to enter Simultaneous mode. (Press [F6]. Use<br />

the up-down arrow keys to select “Simultaneous.” Press the right arrow key<br />

to complete the selection.)<br />

2. Select the input level using the up-down arrow keys. (The default level is 60<br />

dB SPL).<br />

3. Use [F5] to change the source type to “Composite” or “Comp DSP.” The<br />

“Tone” selection will not work in Simultaneous Mode, and the Composite<br />

signal will be substituted automatically.<br />

4. Press [START]. The test will run and the results will be displayed in<br />

Difference graph.