88 FONIX <strong>8000</strong> Hearing Aid Test System<br />

3.8 Custom Functions for F7/F8<br />

The [F7] and [F8] keys can be customized to perform many different functions.<br />

This allows you to quickly adjust settings without going into the local menu.<br />

The following functions are available:<br />

• Display: Switch between Gain and SPL.<br />

• Coupler: Change the coupler type (2cc, CIC).<br />

• Bias Ampl: Change the amplitude of the bias signal (see Section 3.5.3)<br />

• Static: Turn on/off single tone or three frequency average (see Sections<br />

3.4.5 and 3.4.6).<br />

• Batt Size: Change the battery size. This is used in battery current drain<br />

measurements.<br />

• Aid Type: Change the aid type (Standard, AGC, Adaptive AGC). This<br />

affects the delay times used in pure-tone sweeps and Digital Speech.<br />

• Off/Prev: Switch between the signal being in spectrum analysis mode<br />

(source off but measurement running) and at the previous selected amplitude<br />

level.<br />

• Data/Graph: Switch between displaying test results in numerical format<br />

and graphical format.<br />

The F7/F8 selections are set in the local menu. In the Coupler Multicurve<br />

screen, press [MENU] to open the local menu. Change the F7 and F8 selections<br />

using the arrow keys. The default selection for these settings can be changed in<br />

the Setup Menu. See Sections 2.2.1 and 2.2.4 for details.<br />

3.9 Complete List of Menu Items<br />

The function of all items in the local menu of the Coupler screen are described<br />

in this section. Enter the local menu by pressing [MENU] in the Coupler screen.<br />

Selections are made by using [, ] to choose an item and [, ] to change the<br />

setting.<br />

There is a dotted or dashed line in between an item and its selection. If the line<br />

is dotted, the selection will only effect the Coupler screen. If the line is dashed,<br />

the selection is global and will have an effect on other measurement screens.<br />

The local menu is divided into five main categories: Display, Measurement,<br />

Source, Aid and Misc Settings. To skip ahead to the next category, use the<br />

[START] button.<br />

Display Settings<br />

• Data/Graph: The method of data display. Choose between DATA (numerical<br />

data) and GRAPH.