General Operation 41<br />

Figure 2.3.2.5: ANSI weighing of the Composite signal at 70 dB SPL RMS<br />

2.3.2.6 Digital Speech technical details<br />

The digital speech signal works by turning the composite signal on for a set<br />

period of time, taking a measurement, turning the signal off for a random period<br />

of time, and turning the signal back on again. This sequence is repeated until the<br />

test ends.<br />

The signal “on” time takes a minimum of 60 milliseconds. This includes the<br />

time to turn the signal on, take the measurement, and turn the signal off. Some<br />

aids may require an additional delay before the measurement is taken in order<br />

to allow their circuits to settle. This additional delay time can be changed in the<br />

MISC MEAS delay setting that is available in the local menu of the measurement<br />

screens and in the Setup Menu. The MISC MEAS setting is automatically adjusted<br />

according to the aid type selected. This setting is added to the 60 ms minimum<br />

measurement time.<br />

The default settings are:<br />

Linear: 20 ms<br />

AGC: 100 ms<br />

ADAPTIVE: 200 ms<br />

The signal “off” time is random between 100 and 300 milliseconds. This time is<br />

not controllable by the user.