Real-Ear Measurements 173<br />

SPL input level. If the clinician changes the input level to 80 dB SPL, the target<br />

will decease, showing the prescription for that input level. When the unaided<br />

(REUG 1) is selected, the target curve uses the source settings for REAR 3.<br />

In the Real-ear SPL screen, up to three SPL targets are displayed on the screen.<br />

When NAL-NL1 or MOD NAL is the selected fitting rule, all three targets are<br />

automatically displayed. They are labeled L, M and H. The L target is tied to<br />

the input level and speech weighting selected for REAR 2. The M target is tied<br />

to the input level and speech weighting selected for REAR 3. The H target is<br />

tied to the input level and speech weighting selected for REAR 4. That is, if the<br />

clinican changes the input level or speech weighting of REAR 2, 3 or 4, the corresponding<br />

SPL target will also adjust.<br />

If a fitting rule other than NAL-NL1 or MOD NAL is selected, only the M target<br />

is displayed, which is tied to REAR 3. If desired, the clinican can turn on<br />

the display of the other two targets using [F3], but these targets will be linearly<br />

related to each other. That is, unlike with the NAL-NL1 or MOD NAL non-linear<br />

targets, the shape of the linear target will not change as the input level of the<br />

input signal changes.<br />

8.4.7 Using Semi-Auto and Fully-Auto Mode<br />

The Real-ear Insertion Gain and Real-ear SPL screens have basic implementations<br />

of Semi-Automatic and Fully-Automatic test modes. In Semi-Automatic<br />

mode, the analyzer automatically advances to the next measurement curve<br />

when the measurement is stopped. That is, if the user selects REUR 1 and takes<br />

a measurement, the analyzer will automatically advance the curve selection to<br />

REAR 2. This eliminates the need for the clinician to press [F2] and manually<br />

advance to the next curve. Semi-automatic mode is available in both the Realear<br />

Insertion Gain and Real-ear SPL screens.<br />

to activate semi-automatic mode:<br />

1. Enter the Real-ear Insertion Gain or Real-ear SPL screen.<br />

2. Press [MENU] to open the local menu.<br />

3. Select Auto Test under Measurements [, ]. Pressing [START] will jump to<br />

the beginning of the next section in the local menu.<br />

4. Choose Semi-Auto using [, ].<br />

5. Press [EXIT] to return to the real-ear measurement screen. The analyzer will<br />

now be in Semi-Automatic mode.<br />

Fully automatic mode automatically measures REAR 2, REAR 3, REAR 4, and<br />

REAR 5 in quick succession. When REUR 1 is selected, after pressing [START]<br />

to perform the unaided measurement, the user must still press [STOP] to stop<br />

the test. This will automatically advance the curve selection to REAR 2, but no<br />

measurement will be taken until the [START] button is pressed again.