72 FONIX <strong>8000</strong> Hearing Aid Test System<br />

This type of distortion is only apparent when the composite signal source is<br />

used because pure-tone source types do not contain more than one frequency at<br />

the same time.<br />

Figure 3.4.8—Measuring intermodulation distortion<br />

3.4.9 Battery Current Drain<br />

You can measure the battery current drain in the Coupler screen if you use<br />

battery pills (instead of regular batteries) in the hearing aid while performing<br />

your measurements. To turn on the battery current drain measurement in the<br />

Coupler screen:<br />

1. Set up the hearing aid in the sound chamber for testing. You must use a<br />

FONIX battery pill to test for current drain.<br />

2. Close the lid of the sound chamber and latch it.<br />

3. Press [MENU] in the Coupler screen.<br />

4. Use [, ] to select Battery Meas under Misc Settings. (You can use<br />

[START] to skip between groups of selections.)<br />

5. Use [] to turn ON the Battery Meas.<br />

6. Use [] to select Battery Size and use [, ] to select the hearing aid’s battery<br />

size.