Real-Ear Measurements 175<br />
When the reference microphone is disabled, the analyzer uses the information<br />
determined during the leveling process to determine the signal outside the ear<br />
during the measurement. It is very important that the patient remain still after<br />
leveling and that the sound field is undisturbed, particularly if the reference<br />
microphone has been disabled.<br />
to enable/disable the reference microphone:<br />
1. Enter the Real-ear Insertion Gain or Real-ear SPL screen.<br />
2. Press [MENU] to enter the local menu.<br />
3. Select Reference Mic using [, ] under Measurements.<br />
4. Use [, ] to choose ON or OFF.<br />
5. Press [EXIT] to return to the measurement screen.<br />
8.4.10 Viewing numerical curve data<br />
The frequency response curves measured by the <strong>8000</strong> test system can be viewed<br />
in either the standard graphical format or in data format, allowing the clinician to<br />
take a look at the actual numerical values of the measurements.<br />
to view curve data in the real-ear insertion Gain screen:<br />
1. Enter the Real-ear Insertion Gain:<br />
2. Press [MENU] to enter the local menu.<br />
3. Select Data/Graph under Display using [, ].<br />
4. Use [, ] to choose:<br />
• AIDED DATA: To view selected aided curve.<br />
• IG DATA: To view selected insertion gain curve.<br />
• GRAPH: To view in standard graphical (non-numeric) format.<br />
5. Press [EXIT] to return to the Insertion Gain screen. Instead of the measurement<br />
graphs, you will now see tables of numerical data. See Figure 8.4.9.<br />
6. Use [F2] to select the curve you want to view. Depending upon your selection<br />
in Step 3, you will either see the insertion gain or the aided gain results.