Coupler Multicurve 57<br />

MICROPHONE<br />

O-RING<br />

Figure 3.3.1—Direct access coupler (HA-1)<br />

1. Roll the Fun-Tak, provided with<br />

each instrument, into a rod long<br />

enough to go around the transmitting<br />

end of the aid (approximately 2<br />

inches). (Modeling clay can also be<br />

used, but it doesn’t work as well.)<br />

2. Bend the Fun-Tak rod around the<br />

canal of the aid, making the resulting<br />

“donut” flush with the end of<br />

the aid. (Some users choose to seal<br />

the vent opening at this end with a<br />

small amount of Fun-Tak.)<br />

3. Align the sound opening of the aid<br />

to the hole at the conical end of<br />

the coupler. Look through the open<br />

end of the coupler to be sure the<br />

sound opening of the aid is clear of<br />

obstructions and correctly placed.<br />

4. Seal any vent on the aid with a<br />

small kernel of Fun-Tak.