Automated Test Sequences 113<br />

5.2.4 Viewing ANSI 92 Results<br />

1. Family of composite gain curves made at 50, 60, 70, 80 and 90 dB SPL,<br />

2. NSPL90—Maximum RMS output sound pressure level (SPL) produced with<br />

a 90dB RMS speech-weighted noise input SPL signal.<br />

3. Full-On Noise Gain—Maximum RMS gain with a 60dB SPL noise input signal.<br />

(Note: This test may not elicit a true reading of maximum gain for hearing<br />

aids with an onset of non-linear operation below 60dB.)<br />

4. Target Reference Gain—This information is only displayed when the actual<br />

gain is greater than the calculated desired setting of gain. This target is used<br />

to adjust the hearing aid’s actual gain to a level of ±1 dB of the target. The<br />

figure is calculated by adding the 60dB SPL input + 17dB and subtracting it<br />

from the NSPL90 RMS output. (NSPL90 D-77 = desired setting of gain)<br />

5. Actual Reference Gain—The measured amount of gain when the hearing aid<br />

is set to the reference test position.<br />

6. Noise I/O Curve –This demonstrates the compression or limiting of the nonlinear<br />

circuit.<br />

For a more detailed explanation of these terms, consult the ANSI S3.42-1992<br />

standard.<br />

Figure 5.2.4—ANSI 92 test results