108 FONIX <strong>8000</strong> Hearing Aid Test System<br />

9. RTG Measured: Measured reference test gain<br />

10. EQ Inp Noise: Equivalent input noise measurement<br />

11. Resp Limit: The response limit level is determined by taking the three frequency<br />

average of the RESP 50 curve and subtracting 20 dB.<br />

F1 is where the RESP 50 curve crosses the response limit level on the way<br />

up.<br />

F2 is where the RESP 50 curve crosses the response limit level on the way<br />

down.<br />

12. HFA/SPA: Three frequency average chosen in the MENU.<br />

13. TDH: total harmonic distortion measurements and the frequencies and<br />

amplitudes at which they are taken.<br />

14. HFA/SPA-SPLITS: Three-frequency average of the telecoil frequency<br />

response curve.<br />

15. RSETS: The difference between the HFA/SPA of the microphone response<br />

curve and the telecoil response curve.<br />

16. Battery Current measurement<br />

17. I/O curves: the input/output measurements taken of AGC aids. Only selected<br />

frequencies are tested.<br />

18. I/O curve selection box: the key to the I/O graph, listing which frequencies<br />

go with which curve.<br />

19. Attack and release measurements, including the frequencies at which they<br />

were measured.<br />

Test Result Differences with ANSI 96 and ANSI 87<br />

• In ANSI 87 and ANSI 96, the response curve (2) is sometimes measured<br />

at 50 dB SPL<br />

• In ANSI 87, I/O (17) is only measured at 2000 Hz<br />

• In ANSI 87, attack and release (19) is only measured at 2000 Hz<br />

• In ANSI 87, AGC aids are measured at full-on gain for the entire test<br />

sequence<br />

• In ANSI 87, the OSPL90 curve is called SSPL90<br />

• In ANSI 96, RSETS (15) is called STS-SPLITS<br />

• In ANSI 87, the telecoil measurement (3, 14, 15) is much different. There<br />

is only one measurement called “TCOIL 10 mA/m”<br />

To delete the displayed results, press [F6] and use [,] to select DELETE<br />

ALL CRVS. Press [] to perform the deletion.