02.08.2013 Views

FONIX® 8000 - Frye Electronics

FONIX® 8000 - Frye Electronics

FONIX® 8000 - Frye Electronics

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

General Operation 31<br />

I/O MEAS DELAY: The amount of time each subsequent tone in an I/O sweep is presented<br />

before the measurement is made. See Section 2.3.1.2.<br />

RESET FREQ: The frequency the analyzer returns to when RESET is pressed.<br />

AVG FREQS: The frequencies used for averaging in pure-tone measurements. Each<br />

frequency set is represented on the screen by the highest frequency in the set. The<br />

sets are:<br />

HFA (High Frequency Average) 1000, 1600, 2500<br />

SPA (Special Purpose Average) 800, 1250, 2000<br />

SPA 1250, 2000, 3150<br />

SPA 1600, 2500, 4000<br />

SPA 2000, 3150, 5000<br />

IEC–(HAIC) 500, 1000, 2000<br />

Coupler Default Settings<br />

BATTERY MEAS: Status of the battery current measurement. Choose ON or OFF.<br />

BATTERY SIZE: Size of battery used with the hearing aid being tested. Used in battery<br />

current measurements.<br />

BATTERY CHEMISTRY: Type of chemistry of the hearing aid battery. Choices are<br />

ZN-AIR(ZincAir) and SILVER. This choice will almost always be ZN-AIR.<br />

NOISE RED (TONE): Amount of pure-tone noise reduction used in coupler measurements.<br />

See Section 2.3.1.1.<br />

NOISE RED (COMP): Amount of composite noise reduction used in coupler measurements.<br />

See Section 2.3.2.1.<br />

RESET SOURCE: The signal type selected in the Coupler screen when the [RESET]<br />

button is pushed.<br />

DISPLAY: Display of the Coupler Multicurve screen. Choose between Gain and SPL.<br />

CURVE LABEL: Label used for the two available “ears.” Choose between LEFT/<br />

RIGHT and A/B.<br />

GRAPH SCALING: Scaling used when making measurements. NORMAL scales to<br />

the upper-most curve displayed. AUTO scales to the current selected curve.<br />

COUPLER TYPE: The type of coupler being used in making sound chamber measurements.<br />

Selections of CIC and MZ turn on corresponding software correction<br />

factors (see Sections 3.6 and 3.7). A selection of NONE or 2cc does not apply correction<br />

factors to the curve.<br />

STATIC TONE: Status of individual frequency measurement. Choose SINGLE to<br />

emit a static tone in between composite measurements or pure-tone sweeps. Choose<br />

AVG to emit a three frequency average between other measurements. Choose OFF<br />

to turn off this measurement.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!