114 FONIX <strong>8000</strong> Hearing Aid Test System<br />

5.3 IEC<br />

5.2.5 Viewing ANSI measurements in the Coupler screen<br />

It’s possible to copy ANSI response curves to the Coupler screen where they<br />

can be compared to other response curves. To do this:<br />

1. Run the ANSI 92 test sequence.<br />

2. Press [F6].<br />

3. Use [] to select Store Crvs Coup, if necessary.<br />

4. Press [] to store the curves and close the menu. The following is a list of<br />

where the curves are stored in the Coupler screen. Any previous measurements<br />

stored in these curve locations will be overwritten.<br />

• 50 dB SPL—CRV3<br />

• 60 dB SPL—CRV4<br />

• 70 dB SPL—CRV5<br />

• 80 dB SPL—CRV6<br />

• 90 dB SPL—CRV7<br />

5. Press [EXIT] and [F1] to enter the Coupler screen and view the ANSI<br />

response curves.<br />

5.2.6 Explaining the Menu Settings<br />

There are only a couple of menu settings in the local menu of the ANSI 92<br />

screen.<br />

Misc. Settings<br />

• Print Label—Status of the printing label. Choose ON or OFF.<br />

• Printer—Printer used for printouts. Only INTERNAL is available at this<br />

time. See Section 2.4.<br />

The International Electro-technical Commission (IEC) 60118-7 standard in<br />

the international community is analogous to the ANSI standard in the United<br />

States.<br />

There are two versions of the IEC standard in common use: IEC 60118-7 1994<br />

(IEC 94) and IEC 60118-7 2005 (IEC 05). IEC 05 is very similar to the ANSI 03<br />

standard.<br />

Technical note for IEC 94:<br />

Most of the IEC 94 measurements are made with the aid set to reference test<br />

gain. You can either set the aid to the reference test gain calculated by the <strong>8000</strong><br />

analyzer, or you can set it to the reference test gain specified by the manufacturer.