170 FONIX <strong>8000</strong> Hearing Aid Analyzer<br />

to auto-adjust the unaided response:<br />

When using the SPL display in the Insertion Gain screen, or when measuring<br />

in the Real-ear SPL screen (in other words, when the unaided response is displayed<br />

in terms of SPL and not gain), the unaided response includes the input<br />

signal used in the measurement.<br />

In order to directly compare the unaided response to the aided response in an<br />

SPL graph, it is important that the speech weighting and input level of the input<br />

signal are the same for both the unaided and the aided response. For example,<br />

if the unaided response is measured at 65 dB SPL, it is difficult to directly compare<br />

it to the measured aided response at 80 dB SPL when viewing the data in<br />

an SPL graph. However, a direct comparison of the unaided and aided response<br />

can be desirable in ensuring the hearing aid amplifies above the unaided<br />

response. Therefore, the <strong>8000</strong> Hearing Aid Test System has an “REUR Auto<br />

Adjust” feature.<br />

When the REUR Auto Adjust is enabled, the unaided response is automatically<br />

adjusted according to the speech weighting and input level of the selected aided<br />

measurement. This allows the clinician to measure the unaided response at one<br />

level and directly compare it to aided measurement curves measured at another<br />

level.<br />

To enable the REUR Auto Adjust, press [MENU] in the Real-ear Insertion Gain<br />

or Real-ear SPL screen to open the local menu. Use the arrow keys to select<br />

REUR Auto Adjust under Measurements and change its setting to On.<br />

The REUR Auto Adjust is automatically enabled when Unaided is set to Average<br />

in the local menu.<br />

8.4.4 Measuring the Aided Gain in the Insertion Gain Screen<br />

Up to four different aided curves can be measured and displayed together in the<br />

Real-ear Insertion Gain and Real-ear SPL screens. Any measurement made in<br />

one of these screens will automatically be displayed in the other screen.<br />

to measure the aided response:<br />

1. Position the patient in front of the sound field speaker and place the earhook<br />

and reference microphone on the ear. See Sections 8.2.1 and 8.2.2.<br />

2. Insert the probe tube into the patient’s ear canal. See Section 8.2.3.<br />

3. Input the patient’s audiogram to generate a target, if desired. See Section<br />

8.3.<br />

4. Press [EXIT] in the Audiogram Entry screen to return to the Real-ear<br />

Navigation screen.<br />

5. Enter the Insertion Gain screen by using [F3] in the Real-ear Navigation<br />

screen.