Real-Ear Measurements 163<br />

Figure 8.3.5E—Taking an RECD Measurement<br />

8.3.6 Modifying a Target<br />

Once you input an audiogram in the Audiogram Entry screen, a target will automatically<br />

appear on the real-ear measurement screens. You can use the Real-Ear<br />

Target screen to modify a target or to input your own target in dB insertion gain<br />

or dB SPL, edit your client’s HTL and UCL values in dB SPL, and input a custom<br />

real-ear to dial difference (REDD).<br />

1. Enter the Real-Ear Navigation screen by pressing [F2] from the Opening<br />

screen or [EXIT] from a real-ear screen (such as the Audiogram Entry<br />

screen).<br />

2. Press [F2] to enter the Real-Ear Target screen. See Figure 8.3.6. If you have<br />

input an audiogram as described in Section 8.3.1, you should see targets in<br />

both the SPL and gain graphs.<br />

3. Select the desired ear with [F1].<br />

4. Use [F3] to select the desired fitting rule.<br />

5. Use [F2] to select the data you want to edit. The choices are:<br />

• HTL (re): The client’s audiogram in dB SPL<br />

• Target (re): The mid-level target in dB SPL<br />

• UCL (re): The client’s uncomfortable levels in dB SPL