68 FONIX <strong>8000</strong> Hearing Aid Test System<br />

3.4.4 Deleting measurements and settings<br />

Use the [F4] key to permanently delete measurement curves. When you press it,<br />

there are four possible selections:<br />

• Selected Curve: Delete only the selected curve, but keep the curve selections<br />

(usually source type and level).<br />

• Sel Curve & Settings: Delete selected curve and return curve to analyzer<br />

default setting.<br />

• All Curves: Delete all curves, but keep the curve selections.<br />

• All Curves & Settings: Delete all curves and return them all to analyzer<br />

default settings.<br />

To delete a curve:<br />

1. Press [F4].<br />

2. Use [, ] to make the desired deletion type.<br />

3. Press [] to perform the deletion and close the pop-up menu.<br />

3.4.5 Running a Single Frequency Measurement<br />

At times, you may want to find out what the frequency response of the hearing<br />

aid is to a single pure-tone signal. See Figure 3.4.5.<br />

1. Press [MENU] from the Coupler screen.<br />

2. Use [, ] to select Static Tone under Source Settings.<br />

3. Use [, ] to choose SINGLE.<br />

4. Press [EXIT] to return to the Coupler screen. The single tone will automatically<br />

be running.<br />

5. Adjust the frequency using the [, ] buttons. Adjust the amplitude using<br />

the [,] buttons.