220 FONIX <strong>8000</strong> Hearing Aid Test System<br />

Pogo<br />

Prescription of Gain/Output (POGO) for Hearing Aids<br />

by: Geary A. McCandless, PhD,<br />

& Poul Erik Lyregaard, MSc<br />

Hearing Instruments vol 34 #1, 1983<br />

Required insertion gain is calculated as follows:<br />

Frequency (Hz) Insertion Gain (dB)<br />

250 1/2 HTL - 10 dB<br />

500 1/2 HTL - 5 dB<br />

* 750 1/2 HTL - 2.5 dB<br />

1000 1/2 HTL<br />

* 1500 1/2 HTL<br />

2000 1/2 HTL<br />

3000 1/2 HTL<br />

4000 1/2 HTL<br />

* 6000 1/2 HTL<br />

* <strong>8000</strong> 1/2 HTL<br />

Note: Frequencies preceded by an asterisk (*) are<br />

interpolated because the article does not reference them.<br />

BERgER<br />

A Method of Hearing Aid Prescription<br />

by: Kenneth W. Berger, PhD, Eric N. Hagberg, M.A.,<br />

& Robert L. Rane, PhD<br />

Hearing Instruments July 1978<br />

Required insertion gain is calculated as follows:<br />

Frequency (Hz) Insertion Gain (dB)<br />

* 250 1/2 HTL<br />

500 1/2 HTL<br />

* 750 1/2 HTL<br />

1000 1/1.6 HTL<br />

* 1500 1/1.5 HTL<br />

2000 1/1.5 HTL<br />

3000 1/1.7 HTL<br />

4000 1/2 HTL<br />

* 6000 1/2 HTL<br />

* <strong>8000</strong> 1/2 HTL<br />

Note: Frequencies preceded by an asterisk (*) are<br />

interpolated because the article does not reference them.