Real-Ear Measurements 169<br />

5. Enter the Insertion Gain or Real-ear SPL screen by using [F3] or [F4] in the<br />

Real-ear Navigation screen, respectively.<br />

6. Select the desired ear with [F1].<br />

7. Level the sound field speaker. See Section 8.2.4.<br />

8. Use [F2] to select REUR 1. (Press [F2]. Use [, ] to select REUR 1. Press<br />

[] to complete the selection and close the pop-up menu.)<br />

9. Use [F5] to select Composite.<br />

10. Use [, ] to select the desired amplitude for the measurement. We recommend<br />

65 dB SPL.<br />

11. Press [START] to start the measurement.<br />

12 Press [STOP] when the unaided response has stabilized. Curve 1 in the bottom<br />

graph of Figure 8.4.1 shows an example of the unaided response using<br />

the Gain display in the Insertion Gain screen.<br />

13. If desired, use [F1] to switch to the other ear, and repeat steps 7-12.<br />

to choose between Gain and SPL (output):<br />

In the Real-ear Insertion Gain screen, the bottom graph contains the unaided<br />

and aided measurement curves as a function of gain (REUG/REAG) or of output<br />

SPL (REUR/REAR). The “SPL” display includes includes the input signal in the<br />

measurement. The “gain” display subtracts the input signal in the measurement.<br />

1. Enter the Real-ear Insertion Gain screen.<br />

2. Press [MENU] to open the local menu.<br />

3. Select Display using the up-down arrow keys.<br />

4. Use the right arrow key to choose between Gain and SPL.<br />

5. Press [EXIT] to return to the Real-ear Insertion Gain screen. The bottom<br />

graph should now match your selection.<br />

The Real-ear SPL screen cannot display measurements in Gain. All measurements<br />

are displayed in SPL (output). Therefore, the Display selection in the<br />

local menu is disabled.<br />

To use an average unaided response:<br />

It is also possible to display an average unaided response on the measurement<br />

screen instead of a measured response. This allows the clinician to skip the<br />

unaided response measurement. When the unaided response is set to average,<br />

the REAR 2 curve is automatically selected at the first measurement curve<br />

instead of the REUR 1 curve.<br />

To use an average unaided curve, press [MENU] in the Insertion Gain or Realear<br />

SPL screen. Use the arrow keys to set Unaided to Avg. Press [EXIT] to close<br />

the local menu. The average unaided response will now be displayed on the<br />

measurement graph.