Real-Ear Measurements 169<br />
5. Enter the Insertion Gain or Real-ear SPL screen by using [F3] or [F4] in the<br />
Real-ear Navigation screen, respectively.<br />
6. Select the desired ear with [F1].<br />
7. Level the sound field speaker. See Section 8.2.4.<br />
8. Use [F2] to select REUR 1. (Press [F2]. Use [, ] to select REUR 1. Press<br />
[] to complete the selection and close the pop-up menu.)<br />
9. Use [F5] to select Composite.<br />
10. Use [, ] to select the desired amplitude for the measurement. We recommend<br />
65 dB SPL.<br />
11. Press [START] to start the measurement.<br />
12 Press [STOP] when the unaided response has stabilized. Curve 1 in the bottom<br />
graph of Figure 8.4.1 shows an example of the unaided response using<br />
the Gain display in the Insertion Gain screen.<br />
13. If desired, use [F1] to switch to the other ear, and repeat steps 7-12.<br />
to choose between Gain and SPL (output):<br />
In the Real-ear Insertion Gain screen, the bottom graph contains the unaided<br />
and aided measurement curves as a function of gain (REUG/REAG) or of output<br />
SPL (REUR/REAR). The “SPL” display includes includes the input signal in the<br />
measurement. The “gain” display subtracts the input signal in the measurement.<br />
1. Enter the Real-ear Insertion Gain screen.<br />
2. Press [MENU] to open the local menu.<br />
3. Select Display using the up-down arrow keys.<br />
4. Use the right arrow key to choose between Gain and SPL.<br />
5. Press [EXIT] to return to the Real-ear Insertion Gain screen. The bottom<br />
graph should now match your selection.<br />
The Real-ear SPL screen cannot display measurements in Gain. All measurements<br />
are displayed in SPL (output). Therefore, the Display selection in the<br />
local menu is disabled.<br />
To use an average unaided response:<br />
It is also possible to display an average unaided response on the measurement<br />
screen instead of a measured response. This allows the clinician to skip the<br />
unaided response measurement. When the unaided response is set to average,<br />
the REAR 2 curve is automatically selected at the first measurement curve<br />
instead of the REUR 1 curve.<br />
To use an average unaided curve, press [MENU] in the Insertion Gain or Realear<br />
SPL screen. Use the arrow keys to set Unaided to Avg. Press [EXIT] to close<br />
the local menu. The average unaided response will now be displayed on the<br />
measurement graph.