15.03.2018 Views

BAKER HUGHES - Drilling Fluids Reference Manual

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Baker Hughes <strong>Drilling</strong> <strong>Fluids</strong><br />

6. Pour the 30 ml of 0.010 M Potassium Fluoride solution into a plastic beaker.<br />

7. Deliver by pipette or 10cc plastic syringe 3 ml of the Acetic Acid/Potassium Acetate<br />

Buffer into the plastic beaker of 0.010 M Potassium Fluoride.<br />

8. Swirl the beaker contents to ensure proper mixing. Allow the solution to set for five<br />

minutes.<br />

Instrument Calibration, Measuring, and Electrode Care:<br />

Electrode Preparation<br />

The electrode is shipped without filling solution in the reference chamber. To fill from the flipspout<br />

bottle:<br />

1. Lift the spout to a vertical position.<br />

2. Insert the spout into the filling hole in the outer sleeve and add a small amount of filling<br />

solution to the chamber. Tip the electrode to moisten the O-ring at the top and return<br />

electrode to a vertical position.<br />

3. Holding the electrode by the barrel with one hand, use the thumb to push down on the<br />

electrode cap, allowing a few drops of filling solution to drain to wet the inner cone.<br />

4. Release sleeve. If sleeve does not return to its original position immediately, check to see<br />

if the O-ring is moist enough and repeat steps 2-4 until the sleeve has returned to original<br />

position. Add filling solution up to the filling hole.<br />

5. The probe is now ready for use<br />

Add filling solution each day before using electrode. The filling solution level should be at least<br />

one inch above the level of sample in the beaker to ensure a proper flow rate. If the filling<br />

solution is less than one inch above the sample solution level, electrode potentials may be erratic.<br />

Calibration Procedure for Orion Model 720A<br />

1. Fasten the electrode into the appropriate jack on the meter.<br />

2. Turn the instrument on and wait approximately one minute.<br />

3. Press [1ST] and then press [MODE]. Continue to press [MODE] until the LCD reads<br />

“CON” (concentration).<br />

4. Press [1ST], then press [CALIBRATE]. The instrument will display time, date, year, and<br />

when last calibrated. The instrument will then display the number of calibrators be used.<br />

5. Place probe in the 0.0010 (10) standard solution.<br />

6. Enter the # of calibrators to be used, in this case 2 , then press [YES]. The instrument will<br />

start to measure the concentration. A minute will elapse and the instrument will display<br />

“ENTER CONCENTRATION”. Enter (10) and press [YES].<br />

7. Remove probe from the 0.0010 standard solution. Rinse the probe with distilled water and<br />

blot with a paper towel. DO NOT WIPE PROBE!<br />

8. Place probe in the 0.010 (100) standard. After 1 minute, the instrument will display<br />

“ENTER CONCENTRATION”. Enter (100) and press [YES].<br />

9. The instrument will start to show several displays. Monitor the mv reading. It should<br />

be within the range of 55-60. (If the mv reading falls outside this range, the electrode<br />

should be replaced.)<br />

Baker Hughes <strong>Drilling</strong> <strong>Fluids</strong><br />

<strong>Reference</strong> <strong>Manual</strong><br />

Revised 2006 3-91

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!