Baker Hughes <strong>Drilling</strong> <strong>Fluids</strong><br />

Equipment and Reagents<br />

1. pH/mv Meter (Orion or instrument capable of using a fluoride electrode and<br />

having concentration mode capabilities)<br />

2. Combination Fluoride Electrode with BNC connector - Orion 9609BN (currently<br />

supplied with Ion plus A Optimum Results filling solution<br />

3. <strong>Reference</strong> Electrode Filling Solution - Orion Ion plus A Optimum Results filling<br />

solution (Orion 900061) or IONALYZER reference electrode filling solution<br />

(Orion 900001)<br />

4. Plastic Pipette - 0.5 ml<br />

5. Plastic Beakers (2) - 150 ml<br />

6. Plastic Graduated Cylinder(2) - 50 ml<br />

7. Pipette (3 ml) or 10cc plastic syringe<br />

8. Potassium Fluoride Solution - 0.0010 M (10 Standard)<br />

9. Potassium Fluoride Solution - 0.010 M (100 Standard)<br />

10. Acetic Acid/Potassium Acetate Buffer<br />

Standards Preparation for Instrument Calibration<br />

Note:<br />

Graduated cylinders, beakers, and pipettes should be rinsed with distilled water and dried<br />

between each use.<br />

1. Pour the 30 ml of 0.0010 M Potassium Fluoride solution into a plastic beaker.<br />

2. Deliver by pipette or 10cc plastic syringe 3 ml of the Acetic Acid/Potassium<br />

Acetate Buffer into the plastic beaker of 0.0010 M Potassium Fluoride solution.<br />

3. Swirl the beaker contents to ensure proper mixing. Allow the solution to sit for<br />

five minutes.<br />

4. Place 30 ml of 0.010 M (100 Standard) Potassium Fluoride solution into a plastic<br />

graduated cylinder.<br />

5. Pour the 30 ml of 0.010 M Potassium Fluoride solution into a plastic beaker.<br />

6. Deliver by pipette or 10cc plastic syringe 3 ml of the Acetic Acid/Potassium<br />

Acetate Buffer into the plastic beaker of 0.010 M Potassium Fluoride.<br />

7. Swirl the beaker contents to ensure proper mixing. Allow the solution to sit for<br />

five minutes.<br />

Instrument Calibration, Measuring, and Electrode Care<br />

Electrode Preparation<br />

The electrode is shipped without filling solution in the reference chamber. To fill from the flipspout<br />

bottle:<br />

Baker Hughes <strong>Drilling</strong> <strong>Fluids</strong><br />

<strong>Reference</strong> <strong>Manual</strong><br />

Revised 2006 3-101