01.03.2022 Views

CLINICAL LAB SCIENEC

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

CHAPTER 8: PIPETTING AND USE OF GLASSWARE 203

pipette involves modifying the internal tensioning of a spring that governs

the piston’s travel distance. Springs subjected to changing tensioning behave

more smoothly and consistently when they are allowed to enjoy an interval

of rest as they settle into their new configuration. A pipette that is left idle

for at least 1 minute after a volume adjustment will perform more accurately

than one that is used too quickly. This is especially important when calibrating

a pipette.

Pipette Calibration

Pipette calibration, along with other maintenance, is periodically required for

pipettes used in diagnostic testing, such as that performed in a clinical laboratory,

to provide continued accurate, consistent, and repeatable operation and

measurements. The intervals between calibration vary depending on several factors,

which include but are not limited to:

• The skill and training of the operators. Skilled operators tend to operate

the instrument more correctly and make fewer accuracy-robbing mistakes.

• The liquid dispensed by the pipette. Corrosive and volatile liquids tend

to emit vapors, which ascend into the pipette shaft even under proper

operating conditions and may corrode the metal piston and springs, or

the seals and o-rings that provide an airtight seal between the piston and

the surrounding sleeve.

• Proper and careful handling. Pipettes that are frequently dropped, are

subjected to careless handling or horseplay, or are not properly stored in

a vertical position will tend to degrade in accuracy over time.

• The accuracy required by the instrument. Applications requiring maximum

accuracy also demand more frequent calibration. Instruments

used for purely research applications or in educational settings generally

require less frequent calibration.

Under average conditions, most pipettes can be calibrated semiannually

(every 6 months) to provide satisfactory performance. Institutions that are regulated

by the U.S. Food and Drug Administration generally benefit from quarterly

calibration, which is every 3 months. Critical applications may require monthly

service, while research and educational institutions may need only annual service.

These are general guidelines, and any decision on the appropriate calibration

intervals should be made based on knowledge and skill. Also, considerations

of the brand and type of pipette in question (some are more reliable than others)

should be included in the decision to calibrate pipettes. The conditions under

which the pipette is used and the operators themselves are also important factors

to consider.

There are two basic methods of calibration for pipettes. One type of calibration

is accomplished through means of gravimetric analysis. This requires

dispensing samples of distilled water into a receiving vessel placed on a precision

analytical balance. The density of water is a well-known constant, and therefore

the mass of the dispensed sample provides an accurate indication of the volume

that is delivered. Relative humidity, room temperature, and barometric pressure

Copyright 2010 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. Due to electronic rights, some third party content may be suppressed from the eBook and/or eChapter(s).

Editorial review has deemed that any suppressed content does not materially affect the overall learning experience. Cengage Learning reserves the right to remove additional content at any time if subsequent rights restrictions require it.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!