01.03.2022 Views

CLINICAL LAB SCIENEC

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

CHAPTER 10: QUALITY ASSURANCE 245

6. This rule also detects systematic error and is applied both within and across

separate runs. It is violated within the control material when the last 10

values for the same control level are all on the same side of the mean. It is

called a shift if violated across control materials when the last 10 consecutive

values, regardless of control level, are on the same side of the mean.

The number of values considered can be decreased if three or more levels of

control are run.

7. The term Delta checks refers to the difference between a patient’s present

laboratory result and a previous result that exceeds a predefined limit.

Delta checks are done on samples from patients who have repeat analyses

of tests that have relatively stable parameters from day-to-day operations.

These parameters may be in chemistry or hematology predominantly, and

are for patients who have at least a several-day hospitalization period. Delta

checks are investigated by the laboratory internally to rule out mislabeling,

clerical error, or possible analytical error. Therefore, failures discovered by

Delta checks may be preanalytical or systematic. Actions to verify reasons

for a failure may be the repeat of the sample, repeat of the previous sample

from the same patient, or a manual method of determining the veracity of

the results. In addition, a call to the physician or nurse may confirm that a

preanalytical error occurred due to the specimen being drawn from an IV

line, mislabeling of the sample, or a drastic change in the patient’s condition.

Documentation on the patient’s report should indicate what actions were

taken and that the specimen was rerun for verification.

Quality Control for All Types of Tests

Clinical laboratories perform qualitative, semiquantitative, and quantitative

tests on a variety of biological specimens. A qualitative test is defined as one in

which a particular characteristic of the specimen is determined to be either present

or absent and is most often reported as positive or negative. Quantitative

tests determine the amount of a particular substance or property by an instrument,

with the result expressed numerically. Some tests are defined as providing

semiquantitative results. For instance, if a very small amount of protein is found

in a urine specimen, it might be reported positive as a trace amount. CLIA 88

requires that a laboratory must establish and follow an appropriate QC program

for all testing performed by the laboratory.

Establishment of Control Limits for Quantitative Tests

For quantitative methods, laboratories may use either “assayed” or “unassayed”

control materials. Assayed controls are defined as such when the manufacturer

of the materials has established the acceptable limits for the control test results.

Generally, the limits represent a 2 to 3 SD range. If the laboratory is using assayed

controls, the results obtained from testing the control material must be within

the assayed limits or range. If control results fall outside the limits, the laboratory

must take corrective action. Some laboratories will establish over time

a narrower range established by the manufacturer of 2 SDs using the assayed

Copyright 2010 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. Due to electronic rights, some third party content may be suppressed from the eBook and/or eChapter(s).

Editorial review has deemed that any suppressed content does not materially affect the overall learning experience. Cengage Learning reserves the right to remove additional content at any time if subsequent rights restrictions require it.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!