Calibration Calibration Warnings, Precautions, and Notes<br />
5.2.3 Notes<br />
NOTE: Only bacterial filters with a low flow resistance must be<br />
connected to the patient module and/or the patient<br />
connection.<br />
NOTE: Use surgical gloves whenever touching or disassembling<br />
valves or other internal components of the Breathing<br />
System.<br />
NOTE: Ensure that the gas supply of the machine always complies<br />
with the technical specification.<br />
NOTE: The APL Valve and PAW gauge numerics are for reference<br />
only. Calibrated patient airway pressure is available on the<br />
ventilator screen.<br />
NOTE: If the machine should show faults during the initial<br />
calibration or testing, the machine should not be operated<br />
until the fault has been repaired by a qualified service<br />
technician.<br />
NOTE: After servicing, functional, sensor and system tests must be<br />
carried out before clinical use.<br />
NOTE: To accommodate additional monitors and other equipment<br />
the AS3000 offers up to two vertical mounting tracks. Use<br />
of unauthorized mounting accessories is not recommended.<br />
NOTE: Always secure any equipment placed on the AS3000’s top<br />
shelf.<br />
AS3000 Service Manual 0070-10-0683 5 - 3