Check Valve Cleaning Periodic Maintenance<br />

8. Secure rods and install final assembly:<br />

1. Both rods must be equal in height.<br />

Secure rods by tightening the locking nuts.<br />

2. Install an Absorber Gasket and a Screen onto the Absorber Mounting Plate.<br />

3. Insert Absorber Mount Assembly onto guide rods, noting the position of the Absorber<br />

Hose Port and Absorber Hose.<br />

4. Install Absorber Hose onto Absorber Hose Port.<br />

5. Install Absorber Locking Mechanism onto guide rods, noting the silk screening must<br />

face the front of the block. When assembling, put the locking mechanism lever in the<br />

"locked" position.<br />

6. Secure the locking mechanism using the (2) original 4mm hex screws. Install the two<br />

Absorber Canister Assemblies.<br />

7. Measure the distance between the absorber mount assembly plate at four different<br />

points. The difference in length should not be more than 1 mm.<br />

9. Perform tests on the check valve:<br />

1. Perform "Start-Up Tests" on page 6-15 and "Performance Verification" on page 6-27.<br />

6 - 12 0070-10-0683 AS3000 Service Manual