Check Valve Cleaning Periodic Maintenance<br />
8. Secure rods and install final assembly:<br />
1. Both rods must be equal in height.<br />
Secure rods by tightening the locking nuts.<br />
2. Install an Absorber Gasket and a Screen onto the Absorber Mounting Plate.<br />
3. Insert Absorber Mount Assembly onto guide rods, noting the position of the Absorber<br />
Hose Port and Absorber Hose.<br />
4. Install Absorber Hose onto Absorber Hose Port.<br />
5. Install Absorber Locking Mechanism onto guide rods, noting the silk screening must<br />
face the front of the block. When assembling, put the locking mechanism lever in the<br />
"locked" position.<br />
6. Secure the locking mechanism using the (2) original 4mm hex screws. Install the two<br />
Absorber Canister Assemblies.<br />
7. Measure the distance between the absorber mount assembly plate at four different<br />
points. The difference in length should not be more than 1 mm.<br />
9. Perform tests on the check valve:<br />
1. Perform "Start-Up Tests" on page 6-15 and "Performance Verification" on page 6-27.<br />
6 - 12 0070-10-0683 AS3000 Service Manual