Assembly Installation Guide<br />

2.2.11 Monitoring Products - Mounting and Electrical Connection<br />

1. Any monitoring system compatible with the GCX mounting system's swivel head may be<br />

mounted to the AS3000’s arm.<br />

NOTE: Use of other monitors and mounting hardware is the<br />

responsibility of the installer.<br />

2. Always make full use of all mounting fasteners and strap capturing devices when<br />

mounting monitors to the AS3000.<br />

3. After mounting a monitor to the AS3000, connect it to one of the AC outlets located on<br />

the rear of the AS3000.<br />

a. Turn on each monitor one at a time and ensure that the circuit breaker holds without<br />

tripping.<br />

b. Dress each line cord neatly along the side of the anesthesia machine so that it can<br />

not be easily pulled or extend far from the main chassis.<br />

2.2.12 Agent Monitor Waste Gas Scavenging<br />

1. Respiratory gas monitoring products have an exhaust port from which waste gas expels.<br />

The exhaust port on the gas monitor must be connected to the open barbed fitting on the<br />

waste gas scavenger.<br />

2. Ensure that a tight connecting fitting is attached to the rear of the gas monitor. Ensure<br />

that the other end of the same tube has a tight fitting connection attached to the waste<br />

gas scavenger’s (WGS) barbed fitting.<br />

3. Dress the exhaust tubing neatly along the side of the anesthesia machine so that it can<br />

not be easily pulled or extend far from the main chassis.<br />

2.2.13 Oxygen Sensor Calibration<br />

NOTE: See ‘‘Periodic Maintenance Schedule of Service Activities’’<br />

on page 6-2 for when to calibrate the oxygen sensor.<br />

1. Preparing the unit<br />

a. Allow the breathing system to warm up and reach thermal equilibrium<br />

(approximately 30-60 minutes).<br />

b. In Standby mode, press the MENU button. The menu screen will appear.<br />

c. Select Service, then input the password 2010 to enter the Service screen.<br />

d. Select Calibration to enter the Calibration screen.<br />

e. Select Oxygen Sensor to enter the Oxygen Sensor Calibration screen.<br />

2. Calibration<br />

NOTE: Do not shake the O 2 sensor during calibration.<br />

NOTE: During calibration, keep the O 2 sensor in a vertical position,<br />

connector side up, and bottom side exposed to room air;<br />

keep the O 2 sensor near the heated block to minimize the<br />

temperature difference from within the heated block.<br />

2 - 6 0070-10-0683 AS3000 Service Manual