Calibration Calibration Procedures<br />

5.5.6 PEEP Valve<br />

2. 30 cmH 2O Calibration<br />

FIGURE 5-7 Connecting the Y-fitting to the Test Port<br />

a. Set the APL Valve to 30 cmH 2 O.<br />

b. Connect the Y-fitting to the Test Port (see FIGURE 5-7)<br />

c. Connect a Pressure Meter between the Y-fitting and the AS3000 (use an adapter if<br />

necessary).<br />

d. Use the O 2 Flush Valve and O 2 Flow Meter to maintain pressure at 30.0 ±0.2<br />

cmH 2 O on the Pressure Meter.<br />

e. Select Continue to start calibration.<br />

f. Follow the on-screen instructions.<br />

FIGURE 5-8 Disconnect the plastic flaps from the Adult Test Lung<br />

1. Calibration<br />

a. Disconnect the plastic flaps from the Adult Test Lung (P/N 0138-00-0012) and<br />

connect it to the Y-fitting.<br />

b. Set the Fresh Gas flow to 0 mL/min.<br />

c. Press the O2 Flush Valve to fully fill bellows.<br />

d. Select Continue to start calibration.<br />

e. Follow the on-screen instructions.<br />

AS3000 Service Manual 0070-10-0683 5 - 11