Installation Guide Installation Checkout Procedure<br />
5. Check N 2O, O 2 and AIR Lines for leaks.<br />
a. Remove the cylinders from the AS3000.<br />
b. Connect the O 2 , AIR and N 2 O hoses to the gas line inlets.<br />
NOTE: If a specific gas source is not available, skip the<br />
corresponding leak test.<br />
c. Verify O 2 , AIR and N 2 O Line Pressure gauges read the corresponding line pressure.<br />
d. Set the AIR flow controller to minimum.<br />
e. Pinch to occlude the O 2, AIR and N 2O hoses.<br />
f. Disconnect each hose from its line pressure source.<br />
g. Verify that the pressures do not drop more than 2 psi over a time period of 20<br />
seconds.<br />
6. Verify that all available line and cylinder gauges are operational.<br />
a. Connect the O 2 supply hose and remove the AIR and N 2O hoses.<br />
b. Verify the O 2 Line Pressure gauge shows the reading of the line pressure.<br />
c. Connect the AIR supply hose and remove the O 2 and N 2 O hoses.<br />
d. Verify that the AIR Line Pressure gauges show the reading of the line pressure.<br />
e. Connect the O 2 and N 2O hoses and remove the AIR hose.<br />
f. Verify that the N 2 O and O 2 Line Pressure gauge shows the reading of the line<br />
pressure.<br />
g. Attach the cylinders to the AS3000's yokes (O 2 and N 2 O or AIR). In the case<br />
where multiple cylinders of the same gas are required, attach one at a time.<br />
h. Open all attached cylinders including the O 2 cylinder. If a cylinders pressure is less<br />
than 500 psi, replace the respective cylinder.<br />
i. Verify that the cylinder pressure gauges register a pressure for each type of gas<br />
attached.<br />
7. Test O 2 Flush operation with only O 2 connected.<br />
a. Connect the O 2 supply hose and remove the AIR and N 2 O hoses.<br />
b. Attach a calibrated flow meter to the Common Gas Outlet.<br />
c. Push the O 2 Flush button and verify a reading between 35 L/min and 50 L/min at<br />
the flow meter.<br />
NOTE: O 2 flow may be greater than 50 L/min at higher altitudes.<br />
d. Reconnect the O 2 , AIR and N 2 O hoses.<br />
8. Test the O 2 :N 2 O ratio system (applicable to -01 units only).<br />
a. Set all flow control knobs to minimum.<br />
b. Increase the flow of N 2O until the O 2 reaches 1 L/min<br />
c. Verify that the N 2O flow is not higher than 3.7 L/min.<br />
d. Return flow control knobs to minimum.<br />
AS3000 Service Manual 0070-10-0683 2 - 9