Installation Guide Installation Checkout Procedure<br />

5. Check N 2O, O 2 and AIR Lines for leaks.<br />

a. Remove the cylinders from the AS3000.<br />

b. Connect the O 2 , AIR and N 2 O hoses to the gas line inlets.<br />

NOTE: If a specific gas source is not available, skip the<br />

corresponding leak test.<br />

c. Verify O 2 , AIR and N 2 O Line Pressure gauges read the corresponding line pressure.<br />

d. Set the AIR flow controller to minimum.<br />

e. Pinch to occlude the O 2, AIR and N 2O hoses.<br />

f. Disconnect each hose from its line pressure source.<br />

g. Verify that the pressures do not drop more than 2 psi over a time period of 20<br />

seconds.<br />

6. Verify that all available line and cylinder gauges are operational.<br />

a. Connect the O 2 supply hose and remove the AIR and N 2O hoses.<br />

b. Verify the O 2 Line Pressure gauge shows the reading of the line pressure.<br />

c. Connect the AIR supply hose and remove the O 2 and N 2 O hoses.<br />

d. Verify that the AIR Line Pressure gauges show the reading of the line pressure.<br />

e. Connect the O 2 and N 2O hoses and remove the AIR hose.<br />

f. Verify that the N 2 O and O 2 Line Pressure gauge shows the reading of the line<br />

pressure.<br />

g. Attach the cylinders to the AS3000's yokes (O 2 and N 2 O or AIR). In the case<br />

where multiple cylinders of the same gas are required, attach one at a time.<br />

h. Open all attached cylinders including the O 2 cylinder. If a cylinders pressure is less<br />

than 500 psi, replace the respective cylinder.<br />

i. Verify that the cylinder pressure gauges register a pressure for each type of gas<br />

attached.<br />

7. Test O 2 Flush operation with only O 2 connected.<br />

a. Connect the O 2 supply hose and remove the AIR and N 2 O hoses.<br />

b. Attach a calibrated flow meter to the Common Gas Outlet.<br />

c. Push the O 2 Flush button and verify a reading between 35 L/min and 50 L/min at<br />

the flow meter.<br />

NOTE: O 2 flow may be greater than 50 L/min at higher altitudes.<br />

d. Reconnect the O 2 , AIR and N 2 O hoses.<br />

8. Test the O 2 :N 2 O ratio system (applicable to -01 units only).<br />

a. Set all flow control knobs to minimum.<br />

b. Increase the flow of N 2O until the O 2 reaches 1 L/min<br />

c. Verify that the N 2O flow is not higher than 3.7 L/min.<br />

d. Return flow control knobs to minimum.<br />

AS3000 Service Manual 0070-10-0683 2 - 9