Functional Tests Periodic Maintenance<br />

6.8.4.6 Oxygen Sensor Calibration<br />

NOTE: Oxygen Sensor Calibration can be performed in all<br />

ventilation modes.<br />

1. Press the MENU key and then use the Navigator Knob to scroll to the Calibrate<br />

menu tab (see FIGURE 6-22). Select the Start Calibration button.<br />

FIGURE 6-22 Calibrate Menu Tab<br />

2. After the Start Calibration button has been selected, the screen shown in<br />

FIGURE 6-23 will be displayed, instructing the user to remove the oxygen sensor from<br />

the Breathing System and expose it to room air for at least 3 minutes before proceeding.<br />

After at least 3 minutes have passed, select the Next button to initiate the calibration<br />

process. The progress bar shown in FIGURE 6-24 will be displayed.<br />

NOTE: If the system is going to be used during the calibration,<br />

insert the O 2 cell plug into the port from which the oxygen<br />

sensor was removed using a push and turn motion.<br />

6 - 22 0070-10-0683 AS3000 Service Manual