Periodic Maintenance Functional Tests<br />

6.8.5.5 N 2 O Line Pressure Leak Test<br />

1. Remove the N 2 O cylinder from the AS3000.<br />

2. Connect the N 2 O line pressure hose to the line pressure inlet on the AS3000.<br />

3. Pinch the N 2 O line pressure hose to stop N 2 O line flow.<br />

4. Remove the N 2O line pressure hose from the N 2O line source.<br />

5. The N 2 O line pressure gauge should not fall more than 2 psi in 20 seconds.<br />

6. Reconnect the N 2 O line pressure and remove the pinch in the hose.<br />

6.8.5.6 O 2 Line Pressure Leak Test<br />

1. Remove the O 2 cylinder from the AS3000.<br />

2. Connect the O 2 line pressure hose to the line pressure inlet on the AS3000.<br />

3. Pinch the O 2 line pressure hose to stop O 2 line flow.<br />

4. Remove the O 2 line pressure hose from the O 2 line source.<br />

5. The O 2 line pressure gauge should not fall more than 2 psi in 20 seconds.<br />

6. Reconnect the O 2 line pressure and remove the pinch in the hose.<br />

6.8.5.7 AIR Line Pressure Leak Test<br />

1. Remove the AIR cylinder from the AS3000.<br />

2. Connect the AIR line pressure hose to the line pressure inlet on the AS3000.<br />

3. Pinch the AIR line pressure hose to stop AIR line flow.<br />

4. Remove the AIR line pressure hose from the AIR line source.<br />

5. The AIR line pressure gauge should not fall more than 2 psi in 20 seconds.<br />

6. Reconnect the AIR line pressure and remove the pinch in the hose.<br />

6.8.5.8 Cylinder Supply Check Valves Test<br />

1. Close and remove all gas cylinders from the AS3000.<br />

2. Verify the cylinder gauges return to zero.<br />

3. Using the Digital Flow Meter, verify that no gas is escaping from any of the three yoke<br />

connections on the AS3000.<br />

4. Verify there is no external damage to the gas cylinder or yoke pins.<br />

5. Reconnect the cylinder(s) as necessary.<br />